Show simple item record

Design and testing of conductive polysilicon beam leads for use in a high-density biomedical connector

dc.contributor.authorNikles, Stefanen_US
dc.contributor.authorBradley, Robert M.en_US
dc.contributor.authorBledsoe, Sanford C., Jr.en_US
dc.contributor.authorNajafi, Khalilen_US
dc.date.accessioned2006-12-19T19:09:59Z
dc.date.available2006-12-19T19:09:59Z
dc.date.issued2004-07-01en_US
dc.identifier.citationNikles, Stefan; Bradley, Robert; Bledsoe, Sanford; Najafi, Khalil (2004). "Design and testing of conductive polysilicon beam leads for use in a high-density biomedical connector." Journal of Micromechanics and Microengineering. 14(7): 957-968. <http://hdl.handle.net/2027.42/49046>en_US
dc.identifier.issn0960-1317en_US
dc.identifier.urihttps://hdl.handle.net/2027.42/49046
dc.description.abstractA study was conducted to measure and characterize the reliability of polysilicon cantilever beams with electroplated gold pads for use in a high-density biomedical connector. In this design, an array of beams is brought into contact with a corresponding array of 30 µm high gold bumps, forming electrical connection. Analytical computations of multi-layer beams were performed, including the effects of residual stresses. Beam leads of various lengths and widths with electroplated gold contact pads on their ends were tested over 1000 cycles to determine their mechanical reliability, and to measure their contact resistance with gold bumps on a separate substrate. The dimensions of the polysilicon beam that produce the least breakage were determined to be 400 µm long by 125 µm wide. For a beam having a calculated contact force of ∼100 µN, the initial contact resistance was 764 mΩ. After 1000 connect/disconnect cycles, beams of this type had an average final contact resistance of 1.598 Ω. These results demonstrate that very high-density connectors with high mechanical reliability and low-contact resistance can be fabricated.en_US
dc.format.extent3118 bytes
dc.format.extent1666730 bytes
dc.format.mimetypetext/plain
dc.format.mimetypeapplication/pdf
dc.language.isoen_US
dc.publisherIOP Publishing Ltden_US
dc.titleDesign and testing of conductive polysilicon beam leads for use in a high-density biomedical connectoren_US
dc.typeArticleen_US
dc.subject.hlbsecondlevelPhysicsen_US
dc.subject.hlbtoplevelScienceen_US
dc.description.peerreviewedPeer Revieweden_US
dc.contributor.affiliationumCenter for Wireless Integrated Microsystems, University of Michigan, 2609 Draper Road, Ann Arbor, 48109 MI, USA; Department of Biomedical Engineering, University of Michigan, 2200 Bonisteel Blvd., Ann Arbor, 48109 MI, USAen_US
dc.contributor.affiliationumDepartment of Biological & Materials Sciences, School of Dentistry, University of Michigan, 1011 N. University, Ann Arbor, 48109 MI, USAen_US
dc.contributor.affiliationumDepartment of Otolaryngology, University of Michigan, 1500 E. Medical Center Drive, Ann Arbor, 48109 MI, USAen_US
dc.contributor.affiliationumCenter for Wireless Integrated Microsystems, University of Michigan, 2609 Draper Road, Ann Arbor, 48109 MI, USA; Department of Biomedical Engineering, University of Michigan, 2200 Bonisteel Blvd., Ann Arbor, 48109 MI, USAen_US
dc.contributor.affiliationumcampusAnn Arboren_US
dc.description.bitstreamurlhttp://deepblue.lib.umich.edu/bitstream/2027.42/49046/2/jmm4_7_017.pdfen_US
dc.identifier.doihttp://dx.doi.org/10.1088/0960-1317/14/7/017en_US
dc.identifier.sourceJournal of Micromechanics and Microengineering.en_US
dc.owningcollnameInterdisciplinary and Peer-Reviewed


Files in this item

Show simple item record

Remediation of Harmful Language

The University of Michigan Library aims to describe library materials in a way that respects the people and communities who create, use, and are represented in our collections. Report harmful or offensive language in catalog records, finding aids, or elsewhere in our collections anonymously through our metadata feedback form. More information at Remediation of Harmful Language.

Accessibility

If you are unable to use this file in its current format, please select the Contact Us link and we can modify it to make it more accessible to you.