Citation:Kuang, Cuifang; Hong, En; Feng, Qibo; Zhang, Bin; Zhang, Zhifeng (2007). "A novel method to enhance the sensitivity for two-degrees-of-freedom straightness measurement." Measurement Science and Technology. 18(12): 3795-3800. <http://hdl.handle.net/2027.42/58131>
Abstract: A low cost and high precision system which can measure two-degrees-of-freedom straightness simultaneously is proposed based on a laser collimator. In the system, a miniature fiber coupled diode laser generates a collimating beam. The sensitivity of the straightness error measurement is magnified by a factor of 4 compared with the lateral movement of the retro-reflector itself after the multi-reflections. A Renishaw ML10 laser interference measurement system was used to verify the developed system in a series of experiments in the laboratory environment. Experimental results agree with the expected results. The two-dimensional straightness error measurement had linear correlativity close to 1 and the standard one-to-one deviation error was better than 0.32 µm obtained over a travel range of 600 µm.