<TEXT N="BAC2445">
<BODY>
<DIV1>
<P><PB REF="00000001.tif" SEQ="00000001" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="880" N="00000001">
THE  UNIVERSITY  OF  MICHIGAN
COLLEGE OF ENGINEERING
Department of Mechanical Engineering
Final Report
IMPROVED DISCONTINUITY DETECTION USING COMPUTER-AIDED
ULTRASONIC PULSE-ECHO TECHNIQUES
Julian R. Frederick
James A.'Seydel
ORA Project 360361
support d by:
WELDING RESEARCH COUNCIL
PRESSURE VESSEL RESEARCH COMMITTEE
MATERIALS DIVISION
SUBCOMMITTEE ON NONDESTRUCTIVE TESTING
administered through:
OFFICE OF RESEARCH ADMINISTRATION     ANN ARBOR
December 1972



</P>
<P><PB REF="00000002.tif" SEQ="00000002" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="186" N="00000002">
1/-A'lI;.
~/s&lt;9it



</P>
<P><PB REF="00000003.tif" SEQ="00000003" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="863" N="00000003">
TABLE OF CONTENTS
Page
LIST OF ILLUSTRATIONS                                                           iv
ABSTRACT                                                                       vii
Chapter
I.  INTRODUCTION                                                               1
Objective                                                               1
Deconvolution                                                           1
Plan of the Investigation                                               2
II.  THEORETICAL CONSIDERATIONS                                                 4
Model of an Ultrasonic NDT System                                       4
Analysis of the Model                                                   4
Data Acquisition and Processing                                          3
Analysis of the Data Processing System                                 11
Signal-to-Noise Ratio Analysis                                         17
Interpretation of the Analysis                                         21
III.  EXPERIMENTAL PROCEDURE                                                    25
Equipment                                                              23
Test Samples                                                           26
Procedure                                                              29
IV.  RESULTS                                                                   7a
V.  DISCUSSION OF RESULTS                                                     44
Distance Measurement                                                   241 
Resolution                                                             4
Transducer-Independent Display                                         47
SNR Improvement                                                        47
Compensation for a Scattering Medium                                   48
Applications                                                           56
VI.  CONCLUSIONS                                                               61
VII.  FUTURE WORK                                                               (7
REFERENCES                                                                      64
iii



</P>
<P><PB REF="00000004.tif" SEQ="00000004" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="864" N="00000004">
LIST OF ILLUSTRATIONS
T-lable                                                                          Page
I.  Comparison Between Ultrasonic and Micrometer Distance Measurements                                                                      30
II.  Maximmr, Resolution Expected for the Computer Processed Data in
Figures 9-15                                                               46
Figure
1.  Mathematical model of a. pulse-echo ultrasonic nondestructive
testing system.                                                             5
2.  Loop impulse response of four different transducers.                         6,.  Power spectrum  of the 5-MHz transducer in Figure 2c.                        9
4a.  Graph of a discontinuity delimited by b(t-T1) and 5(t-T2).                  15
4b.  Fraph of s(t) as derived from the g(t) in Figure 4a.                        15
4c.  Graph of a discontinuity delimited by b(t-T1) and b(t-T2).                  16
4d.  Grapnh of s(t) as derived from the g(t) in Figure 4c.                       16
5a.  Model of a conventional pulse-echo ultrasonic system with output fl(t) combined with additive noise n1(t) and inserted into
a rectangular bandpass filter Q(f) designed to optimize the SNR.           19
-b.  Model of the data processing system.                                       19
6.  Schematic diagram of the complete ultrasonic and data processing systems.                                                               24
7a.  Targets B, C, and D consist of a O.508-mm and a 0.787-mm step
in aluminum.                                                               27
7b.  Targets E, F, and G consist of three 1.6-mm diameter holes
drilled in aluminum.                                                       27
8.  Power spectrum  of the multiple reflections in a 26.1l-mm thick
aluminum block.                                                            31
iv



</P>
<P><PB REF="00000005.tif" SEQ="00000005" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="877" N="00000005">
LIST OF ILLUSTRATIONS (Continued)
Figure                                                                         Page
9a.  RF "A" scan presentation of the three-step and reference echoes
using the 10-MHz SFZ transducer in position 2.                           34
9b. Computer drawn graph of the tnree-step target obtained by processing the data shown in Figure 9a.                                     34
10a.  RF "A" scan presentation of the three-step and reference echoes
using the 10-MHz SCJ transducer inl position 2.                         35
lOb.  Computer drawn graph of the three-step target obtained by processing the data shown in Figure 10a.                                    35
11a. RF "'A" scan presentation of the three-step and reference echoes
using the 5-MHz lead metaniobate transducer in position i.               36
l1b.  Computer drawn graph of the three-step target obtained by processing the data shown in Figure lia.                                    56
12a.  RF "A" scan presentation of the three-step and reference echoes
using the 10-MHz SCJ transducer in position 5.                           57
12b.  Computer drawn graph of the three-step target obtained by processing the data shown in Figure 12a.                                    37
13a.  RF "A" scan presentation of the three-step and reference echoes
using a 5-MHz lead metaniobate transducer in position 4.                 38
13b.  Computer drawn graph of the three-step target obtained by processing the data shown in Figure 13a.                                    38
14a.  RF "A" scan presentation of the three-hole and reference echoes
using a 5-MHz lead metaniobate transducer.                               39
14b.  Computer drawn graph of the three-step target obtained by processing the data shown in Figure 14a.                                    39
15a.  RF "A" scan presentation of the tnree-hole and reference ecnoes
using the 2.25-MHz SFZ transducer.                                       40
15b.  Computer drawn graph of the three-hole target obtained by processing the data shown in Figure 15a.                                    40
v



</P>
<P><PB REF="00000006.tif" SEQ="00000006" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="860" N="00000006">
LIST OF ILLUSTRATIONS (Concluded)
Figure                                                                         Page
lsa.  RF!?A' scan presentation of tne target and reference echoes from
a cast iron sample using the 2.27-MHz SFZ transducer.                    41
16b.  Computer drawn graph of cast iror target obtained by processing
thle da ta shown in Figure 16a.                                          41
17a.  RF "A" scan presentation of the three-step and reference echoes
using the 10-MHz SCJ transducer.                                         42
17b.  Computer drawn graph of the three-step target obtained by processing the data shown in Figure 17a.                                    42
lda.  RF "'A" scan presentation of the target and reference echoes from
the plastic block using a 5-MHz lead metaniobate transducer.             45
18b.  Computer drawn graph of the milled step in the plastic block
obtained from the data in Figure lSa.                                    43
19.  Rectified video display of the three-hole target in Figure Ofb
using the 5-MjHz transducer in Figure 2c.                                47
20.  Computer simulation of the effect of scattering on a single
target with dl- d = 0.0 mm.                                              50
21.  Computer simulation of the effect of scattering on a single
target with d1 - d = -12.7 mmi.                                          51
22.  Computer simulation of the effect of scattering on a single
target with d1 - d = -25.4 mm.                                           52
23.  Computer simulation of the effect of scattering on a single
target with d1- d = -50.8 mm.                                             5
24.  Computer simulation of the effect of scattering on a single
target with dl - d = +127 mm.                                            54
25.  Computer simulation of the effect of scattering on a single
target.                                                                   5
26.  Schematic diagram  of a phased array scanning system.                     58
vi



</P>
<P><PB REF="00000007.tif" SEQ="00000007" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="879" N="00000007">
ABSTRACT
The purpose of this project has been to investigate means for obtaining
improved characterization of the size, shape, and location of subsurface
discontinuities in metals.  This has been done by applying computerized data
processing techniques to the signal obtained in conventional ultrasonic pulseecho systems. The principle benefits are improved signal-to-noise ratio, and
resolution.
The received ultrasonic pulse from a discontinuity is combined with a
reference functi on in order to preserve both the phase and amplitude of the
received pulse. Processing is performed on the power spectrum of the resultant
signal.  The reference function is also used to compensate for some of the
frequency-dependent scattering caused by the material microstructure. During
data acquisition, analog signal averaging improves the signal-to-noise ratio
(SMTR).  After data acquisition, a digital deconvolution procedure improves
the transducer longitudinal resolution.
An analysis of a mathematical model of the combined ultrasonic and data
processing system shows that the system provides an order of magnitude increase
in the longitudinal resolution and a 37 dB increase in the signal-to-noise
ratio over that expected from a conventional pulse-echo system.  Experimental
data demonstrate a factor of five improvement in resolution and a 28 dB
increase in the SNR.  The results also show that the computer processed data
are virtually independent of the transducer used in the ultrasonic system.
A theoretical analysis of the model also shows that the final output format is
ideally suited to the formation of a synthetic transducer array designed to
increase the lateral resolution.
Possible applications of this data processing system to current problems
in nondestructive testing are discussed. The testing of materials with large
amounts of internal scattering and frequency dependent attenuation appears to
be particularly promising. Also, the enhanced resolution gained by the system
can be used to reveal previously undetectable discontinuities in thin plates
or those that are near a reflecting surface such as in immersion testing.
vii



</P>
<P><PB REF="00000008.tif" SEQ="00000008" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="877" N="00000008">
I.  INTRODUC TION
BJEC T IV E
The objective of the investigation has been to develop an improved method
for determining the size, shape, and location of a discontinuity using ultrasonic
pulse-echo technique.  This has been done by combining the echo from a discontinuity with a uniform reference echo, as is done in holography,  ild obtaining
their combined power spectrum. The use of a reference echo preserves the
phase and amplitude of the echo from the discontinuity.  The use of a reference
echo also compensates for some of the frequency-dependent scattering caused
by the microstructure of' the material.  During data acquisition, analog signal
averaging improves the signal-to-noise ratio (SNrR). After data acquisition,
a digital deconvolution procedure  improves the transducer longitudinal resolution.
The final output format is ideally suited to the formation of a synthetic
transducer array (using a single scanning transducer) which woul( provide
increased lateral resolution.  Improvements in SNR and longitudinal resolution
have been demonstrated in the present irnvestigation.  The extension of the
work to a synthetic array is a logical next step, but it has not been attempted
in this investigation.
The results of the investigation show that the computer-processed data
are virtually independent of the characteristics of the transducer used in the
ultrasonic system.  This is an important property when data taken over a
period of years are to be analyzed for changes in the number and size of the
discontinuities being monitored.
DEC ONVOLUTI ON
One of the problems in ultrasonic NDT is how to optimize trade-off between
resolution and sensitivity. High resolution requires short-time pulses, and
short-time pulses mean a reduced sensitivity. Sokolov (1941) proposed a
frequency modulation technique which could circumvent this difficulty by
forming pulses with a large time-band-width product, but the technique was
superceded by the pulse-echo method and has never played an important role in
subsequent developments.
When electronic pulses shorter than a microsecond are applied to a transducer, the output pulse shape is primarily determined by the high Q of the
piezoelectric transducer.  The transducer Q can be reduced by mechanical
damping, but again this has a detrimental effect on sensitivity. Deconvolution
of the transducer response is a possible answer to this problem that has been
1



</P>
<P><PB REF="00000009.tif" SEQ="00000009" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="886" N="00000009">
thoroughly discussed in other related fields but never rigorously pursued in
ultrasonics.  Deconvolution will remove the relatively slow time response of
the transducer, and while it normally reduces the system sensitivity by
increasing the noise, sensitivity can be retained, and even increased, by
signal averaging before the deconvolution step. Thus, more resolution and
more sensitivity can be attained simultaneously by a sequence of signal
averaging and deconvolution.
PLAIT OF THE INVESTIGATION
An ultrasonic pulse-echo NDT system is proposed in which improved
discontinuity characterization would be obtained by means of a high resolution,
incremental transducer scan over the test surface at a position above the
suspected discontinuity. The information returned by the target echoes at
each point in this scan would be digitized and stored in a computer. The
digital information from the complete collection of these discrete scanning
points would be manipulated and processed into a form suitable for viewing
the defect geometry on a CRT display.
Since resolution improvement is a key part of such a scanning system the
present effort has been directed toward the development of a computer data
processing system designed to improve both the longitudinal and lateral
resolution. The data processing system divides logically into two parts:
data acquisition and data processing.  Due to the high frequencies involved,
it is not possible to acquire data with the desired accuracy from the usual
time representation of the ultrasonic echoes.  However, sufficiently accurate
data can be acquired from the power spectrum of the ultrasonic echoes, and a
system designed to do this will be described and analyzed.  Analog signal
averaging is a, necessary part of this data acquisition system, and it is
implem-rented immediately before the actual analog-to-digital conversion.
Once the data are in digital form, the piezoelectric trnasducer response
is deconvolved from the complete system response.  This deconvolution procedure
has a. dual benefit. First, it improves the longitudinal resolution of the
ultrasonic system.  This is the primary goal.  Second, it makes the output
display imdependent of any long-term variations in transducer response due to
aging or replacement. With the deconvolved data from many points stored in
the computer, it can process the data to form an ultrasonic array designed to
improve the lateral resolution. Thus, the data processing system can improve
both the longitudinal and the lateral resolution.
Experimental evidence is presented to confirm the feasibility of longitudinal resolution enhancement, SNR improvement, and transducer independence.
The possibility of an ultrasonic array derives naturally from the final form
of the computer processed data, but no experimental data designed to show its
feasibility are presented. The relationship of these three improvements to
2



</P>
<P><PB REF="00000010.tif" SEQ="00000010" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="892" N="00000010">
both the proposed new scanning system and the conventional pulse-echo NDT
system are discussed in detail.
3



</P>
<P><PB REF="00000011.tif" SEQ="00000011" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="867" N="00000011">
II.  THEORETICAL CONS IDERATIONS
MODEL OF AN ULTRASONIC NDiT SYSTEM
A model for the pulse-echo ultrasonic system which has been developed is
shown in Figure 1.  Each box is a linear, time-invariant system that represents
a particular process in the ultrasonic system.  All impulse responses are shown
as functions of time.  In situations where distance is the natural independent
variable, it is transformed into a time variable by the formula t = J/v.  The
pulser provides the input function, and the output function is the electrical
signal representing the target echos. Each box is assigned an impulse response
and a. frequency response as indicated in Figure 1.
A digital data acquisition and processing system accurately samples and
processes the output of the model. This system consists of a wave analyzer, a
low-pass filter, an analog-to-digital converter, a digital deconvoluition, and
an inverse Fourier transform.  After the data are digitized, the model of the
ultrasonic system is enlarged to include the digital data acquisition processing system, and again the response of this larger model to an arbitrary input
and to band-limited Gaussian noise is determined.  A complete theoretical development and justification of the model has been developed by Seydel (Seydel,
1972).
The model is presented in this manner in order to emphasize the fact that
a. conventional ultrasonic system measures the reflectivity function, g(t), only
after it passes through a piezoelectric transducer that modifies and distorts
thle signal.  For example, Figure 2 shows that in a conventional system, a
single discontinuity (the back surface) is represented by a different indication for each different transducer.  As the following analysis demonstrates,
these transducer-induced modifications can be removed, and the reflectivity
function can be directly measured.
ANALYSIS OF THE MODEL
The system output function is now derived using the following results
from linear systems theory (Kaplan, 1962).
1. The system output is equal to the convolution of the system input
with the system impulse response.  The convolution of the two functions a(t) and b(t) is denoted by a(t) * b(t) and is defined to be
the following:
00
a(t) * b(t)  =   f a(t') b(t-t')dt'                                (1)
-00o
4



</P>
<P><PB REF="00000012.tif" SEQ="00000012" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="837" N="00000012">
IMPULSE RESPONSE                            FREQUENCY RESPONSE
p (t)~   g PULSER                                P (f)
h (t)              TRANSDUCER i                  H (f
TEST
m(t)                  MEDIUM                     M (f)
g (t)           DISCONTINUITY                    G (f)
TEST
m(t)                  MEDIUM                     M (f)
h (t)                 _    CER             I   H (f)
Figure 1.  Mathematical model of a pulse-echo ultrasonic nondestructive
testing system. Each physical function is represented by a linear, timeinvariant system.



</P>
<P><PB REF="00000013.tif" SEQ="00000013" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="868" N="00000013">
2a.  2.25 MHz, 13-mm diameter                    2b.  10 MHz, 6.4-mm diameter
transducer.                                      transducer.
2c.  5 MHz, 13 mm-,diameter                      2d.  10 MHz, 6.4-mm diameter
transducer.                                      transduc er.
Figure 2. Loop impulse response of four different transducers. Time scale in 2a is compressed by
a factor of 2.5 compared to the other figures.



</P>
<P><PB REF="00000014.tif" SEQ="00000014" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="867" N="00000014">
2.  The output frequency response is equal to the input frequency response
multiplied by the system frequency response.
5.  The system impulse response and the system frequency response are a
Fourier transform pair.  The Fourier transform of a. lower case time
function, a(t), will be denotea by an upper case frequency function,
A(f), where f is the frequency.
Using the first property, the output-, f(t), of the ultrasonic system  is
written as
f(t)  =  n(t) * m(t) * g(t) * m(t) * h(t)                               (2)
or using properties 2 and 3,
F(f)  =  H2(f) M(f) G(f)                                                ()
Equations (2) and (3) describe the cumulative effect of the transdducer, the
media,, and the discontinuity upon the final output.
Given F(f), each of the terms in equation (5) can be measured by properly
arranging the experimental conditions.  For instance, H(f) can be folnd by
insornifying only a single-point target in a medium with no frequency-dependent
attenuation.  In this case, G(f) = 1, M(f) = 1, and F(f) = H2(f).:,lis result
can then be used to find M(f) by insonifying a, single-point target in a, medium
with frequency-dependent attenuation.  Then F(f) = H2(f) M2(f) and M(f) is
found by dividing this formula by the previously measured H2(f).  As mentioned
earlier, measuring G(f) directly is one of the major goals of this researcl
effort, and a, method for accomplishing tilis task is presented below.
It must be emphasized that H(f) and M(f) are not constant properties of
any particular transducer or medium.  Since the coupling conditions between
the transducer and the medium have been implicitly included in the transducer
response, H(f) can be different for the same transducer in contact with different media,.  Also, M(f) is a, function of the distance of travel of' the
acoustic pulse. This fact becomes important in subsequent derivations when a,
different form of M(f) must be assigned to the defect and reference functionls,
respectively.
Measurement of H(f) and M(f) is complicated by tae fact that, in most
cases, F(f) is not directly measurable.  Almost all spectrum  analyzers (including all high-frequency spectrum  analyzers) measure the power spectrum,
S(f), where S(f) = F(f) F+(f) and F+(f) is the complex conjugate of F(f).  A
method circumventing this difficulty is considered below.
7



</P>
<P><PB REF="00000015.tif" SEQ="00000015" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="880" N="00000015">
The output of the ultrasonic system, f(t), is primarily determined by the
transducer loop response, h(t) * h(t).  If this function is applied to the input of a spectrum analyzer, the output if [H(f) H+(f)]2.  A graph of
[H(f) H+(f)]2 for a high resolution, 5 MHz transducer is shown in Figure 3.
The ordinate in this figure is approximately logarithmic.  Figure 3 demonstrates that the 5 MHz transducer produces power not only at its center frequency but also in a band of frequencies ranging from about 2 MHz to 8 MHz.
DATA ACQUISITION AND PROCESSING
At this point, it is necessary to consider some characteristics of the
digital system that processes f(t).  The most important part of the digital
system is the analog-to-digital converter (ADC) that transforms tne analog
waveform into a set of digital data. The electrical characteristics of the
ADC are largely determined by the upper frequency limit of f(t).  The curve
shown in Figure 5 will be considered as the typical frequency response of an
ultrasonic transducer. In evaluating the ADC, both the time and amplitude
accuracy are specified as 8 bits. The reason for this tight specification is
made clear later.
An ADC contains two basic systems:  a sample-and-hold amplifier preceding
an analog-to-digital encoder.  The sample-and-nold module is necessary because
most high-speed encoders require a constant input signal during the conversion
period. The sample-and-hold module is characterized by two time constants:
the acquisition time and the aperture time.  The acquisition time is defined
as the time required for the output to go from its most negative cenrIdition to
its mrcost positive condition. The acquisition time represents the maximum time
for the output to reacquire and to track +the input.  The aperture time is defined as the time bracket within which the output stops tracking and holds the
input voltage. The actual sampling time within this bracket is a random process. The analog-to-digital encoder is characterized by the time required to
produce a digital representation of the input signal to the necessary accuracy
(usually stated in bits).  More accuracy typically implies a longer conversion
time.
The ADC is operated by sending a timing pulse to the "sample" input of
the sample-and-hold module.  After the acquisition time has elapsed and the
output is tracking the input to within the required accuracy, a timing pulse
is sent to the "hold" input.  The sample-and-hold module then holds this voltage while the digital encoder is activated, and, after the full conversion
time has elapsed, a digital sample is available for storage.  The complete
process is reinitiated by a timing pulse to the "sample" input.
There are two possible techniques for using an ADC  to digitize f(t).  The
first technique uses what is referred to as a free-running ADC to completely
digitize f(t) during one operation of the pulser unit. The free-running ADC



</P>
<P><PB REF="00000016.tif" SEQ="00000016" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="834" N="00000016">
I-     I    I       I,.,       I,I                I 
0.0   1.0    2.0   3.0    4.0   5.0   6.0    7.0    8.0   9.0   10.0
FREQUENCY, MHz
Figure 3. Power spectrum of the 5-MHz transducer in Figure 2c.
The ordinate is approximately logarithmic.



</P>
<P><PB REF="00000017.tif" SEQ="00000017" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="874" N="00000017">
samples and digitizes the waveform under the control of a high-frequency clock.
During each clock pulse, a portion of the analog waveform is sampled, digitally
encoded, and stored.  Since the typical analog waveform has an upper frequency
limit of 8 YHz, the clock frequency must be at least 16 MHz in order to satisfy
the \Nyquist criterion.  To meet the specification of 8-bit accuracy for time
and for amplitude at a i6-MHz clock frequency implies an aperture time of 0.5
ns, an acquisition time of 50 ns, and an 8-bit conversion time of 30 ns. Every
one of these figures is from one to two orders of magnitude beyond the present
state-of-the-art.
The second technique uses the fact that f(t) is a repetitive waveform.
Under these conditions, the ADC can be programmed to slide the sample-and-hold
aperture over f(t) while taking only one digital sample per pulse.  Since the
ultrasonic PRF is about 1 kHz, botn the acquisition time and the conversion
timre can now be increased to around 0.5 ms, and these figures are well within
the state-of-the-art.  Unfortunately, the necessary aperture time is still 0.5
us, and this figure is about two orders of magnitude beyond presently avw ilable
instruments.  If new technology can reduce the aperture time to 0.5 ns or less,
this sampling technique would become a very attractive method of data acquisit ion.
Since neither method of directly digitizing f(t) is feasible, it is necessary to consider a, novel approach to this problem.  This new method employs a
systern that acquires the necessary data from the power spectrum of f(t) by
using a wave analyzer.  This approach has several advantages.  There are many
commercially available wave analyzers that make use of recent advances in
phase-locked loops (PLL). A PLL allows the wave analyzer input frequency to
be tiuned with a stability and accuracy determined only by the stability of a
single reference oscillator, even though there may be as many as three of four
different local oscillators in the instrument.  Since this same reference oscillator can be used to control the ultrasonic PRF, the wave analyzer can be
servolocked to the ultrasonic system to provide drift-free spectral measurements.  With a, reference oscillator stability of 1 part in 106 per day and
drift-free amplitude measurements, it is very easy to meet the 8-bit specification on frequency and amplitude accuracy.  The excellent frequency stability
also makes it possible to use a 200-Hz input bandwidth even when measuring 20iiHz signals.  Such a, narrow bandwidth is very important in reducing the omnipresent noise. The narrow bandwidth also imposes a relatively long settling
tiime that is advantageous for inexpensive data conversion and storage.
The Fourier transform of a pulsed signal can be represented by a comb
function whose amplitude envelope is the Fourier transform of a single pulse.
For signals of infinite duration, this is equivalent to the Fourier series
representation of f(t).  The comb spacing is equal to the ultrasonic pulse
rate frequency.  Spectral measurements are made by tuning the wave analyzer to
one of these comb spikes and measuring the power. The power readings are converted into a digital form by measuring the detector voltage with a digital
voltmeter. This process is repeated for each comb spike.
10



</P>
<P><PB REF="00000018.tif" SEQ="00000018" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="875" N="00000018">
The power spectrum has the singular disadvantage of not preserving the
phase information present in f(t). If F+(f) is the complex conjugate of the
Fourier transform of f(t), then the power spectrum of f(t) is denoted by
F(f) F+(f), and the wave analyzer output is
+~f)p[Fr)  im(f)                 i ](f)     =         2
F(f) F (f)  =  [IF(f)le  (] JF(f)e    f]  =   F(f)l                   (4)
All the phase information contained in ~(f) is no longer present at the output.
Unless stated otherwise, it will be assumed that all physical outp tIs are
complex-valued functions of a real variable (time or frequency).
The phase information in F(f) can be preserved by adding a reference functiot to f(t) prior to its insertion into) the wave analyzer.  The manner in
which the phase information is preserved is shown by the example below.  If a.
delta function is added to the normal ultrasonic system output, the final output is f(t) + 5(t). The Fourier transform of this output is F(f) + 1, and its
power spectrum is IF(f) 12 + 1 + F+(f) + F(f).  If the various terms in this
sum can be separated in some manner, then F(f) can be recovered witn both
phase and amplitude information intact. It is important to note that recovering
F(f) is equivalent to recovering f(t) since they are uniquely related by a
Fourier transform.
The particular phase reference function used can be chosen as a matter of
convenience. If an arbitrary reference function, r(t), is used, then F(f) is
multiplied by R (f).  In the example above, this modification would be undesirable because F(f) alone is required and no more processing is contemplated.
But if additional processing is desirable, then r(t) can be chosen so as to
simplify the subsequent systems.  For example, in the deconvolution system described below, r(t) is chosen so that the deconvolution filter is an amplitude
only filter. Since such a deconvolution filter has no phase reversals or axis
crossings to complicate matters, this choice of r(t) results in a considerable
simplification.
ANALYSIS OF THE DATA PROCESSING SYSTEM
Since the high Q impulse response of the piezoelectric transducer is primarily responsible for the slow time response of the ultrasonic system, a deconvolution system that compensates for the transducer impulse response, h(t),
is very attractive. This is accomplished in the following manner. Assume
that an arbitrary stress function, a(t), is applied to the transducer input,
and that the transducer output voltage is b(t).  Then
00
b(t)  =  S a(t') h(t-t')dt'                                           (5)
-00
11



</P>
<P><PB REF="00000019.tif" SEQ="00000019" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="870" N="00000019">
Since h(t) can be measured, and b(t) is known, this integral equation can be
solved for a(t). A formal solution to the deconvolution problem is quickly
achieved by taking the Fourier transform of both sides of the convolution integral.  Then B(f) = A(f) H(f) and A(f) = B(f)/H(f).  Now, a(t) is found by
taking the inverse Fourier transform of A(f).  If H'(f) is defined as l/H(f),
then A(f) = B(f) H'(f) and the deconvolution process becomes an additional
linear system whose input is b(t) and whose output is a(t).  Since the frequency response of the deconvolution process is the inverse of the transducer
frequency response, the deconvolution process is frequently called the "inverse
filter" for h(t), in an analogy with electrical filters.
The application of a digital deconvolution procedure to the complete
ultrasonic system is a fairly simple matter since the data acquisition system
uescribed above already provides digital spectral data.  The first step in
forming the inverse filter is to measure h(t) and b(t), or equivalently H(f)
and B(f).  This process is illustrated by computing the output of the ultrasonic system when two different types of target reflectivity functions are
assumed.
Example 1. Assume that the target is a point reflector located a time T
away from the transducer.  Then g(t) = b(t-T) and the output, fl(t), is
f (t)  =  h(t) * m(t) * 6(t-T) * m(t) * h(t)
=  h(t) * m(t) * m(t-T) * h(t)                                   (6)
The Fiourier transform of f](t) is
Fl)         2(f) M(f) =  H(f)M2(f)  e2                                   (7)
and1 the power spectrum, S(f), of fl(t) is
SL(f)  =  [H(f) H+(f)]2 [M(f) M*(f)]2                                   (8)
Example 2.  Assume that g(t) is any arbitrary reflectivity function
localized in an area, where the media, response is denoted by ml(t), and that a,
reference function is derived from a point target located a, time T from the
transducer.  Then
12



</P>
<P><PB REF="00000020.tif" SEQ="00000020" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="789" N="00000020">
f2(t)  =  [h(t) * ml(t) * g(t) * ml(t) * h(t)]
+ [h(t) * m(t) * m(t-T) * h(t)]                              (9)
and
F2(f)  =  H2(f) Ml(f) G(f) + H2(f) M(f) ei 2 f
-  H (     f) M(f) L   G(f) + e-i2J fT
2  2(M (f)
H (f)M2(f) LM 2   G(f) ( + ei 2    J                        (10)
where
M2(f)
M2(f)       2(                                                        (11)
So(f)  =  [H(f) H+(f)]  ) Mf) M (f)]2
2M2   (f)]  [M-f
[M(f) M(f) G(f) G (f) + 1 + e
G (f) Mo(f) + e          G(f) M(f)                          (12)
The deconvolution procedure now becomes obvious.  The first two bracketed
terms in S2(f) are exactly equal to Sl(f). Thus
S2(f)  =  Sl(f)  M(f) M2(f) G(f) G(f) + 1 + e
G (f) M2(f) + e        M2(f) G(f)                  (13)
For the moment, consider the case where the scattering due to the medium is
negligible at all points within the test block. Thus M(f) = Ml(f) = M2(f)  1.
A thorough treatment of the effects of the medium will be postponed until Section V. Then by defining S(f) = S2(f)/Sl(f)
13



</P>
<P><PB REF="00000021.tif" SEQ="00000021" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="855" N="00000021">
S(f)  =  G(f) G (f) + 1 + e              G(f) + e        f  T G(f)      (14)
and taking the inverse Fourier transform of S(f) gives
s(t)  =   r'(t) + g(-t +T) + g(t-T)                                     (15)
There g'(t) is the inverse Fourier transform of [G(f) G+(f) + 1].  Equation
(15) presents tne form of the output from the complete ultrasonic and data
processing system.  As promised, it is a direct measure of the reflectivity
function, including both the phase and the amplitude.
First, it, is necessary to establish that g(t) can be separated from the
other terms in s(t).  Since g(t) is generally localized in a small region of
the t-axis, let this region be delimited by two delta functions, 6(t-T1) and
t(t-T2).  Then
-i2itf(T1-T )    i2t f(T1-T2)    i2irf(T1-T)
S(f)  =  3 + e                 + e                + e
i 2A f(T2-T)    -i 2i f(T -T)    -i 2t f(T2-T)
+ e              + e               +e                       (16)
and
s(t)  =  36(t) + b(t-T +T2) + b(t+T1-T2) + 6(t+T1-T)
+ b(t+T2-T) -+ b(t-Tl+T) + 5(t-T2+T)                   (17)
The second anld third terms in s(t) are referred to as the cross products, and
the fourth through seventh terms are g(t) and its reflection througn the t = 0
axis.  The reference function can come either before or after the target g(t).
If the reference function comes after the targets, as shown in Figure 4a, then
s(t) is as graphed in Figure 4b.  This is the case for all the data presented
in Section IV.  Since S(f) is real, s(t) is symmetrical about the t = 0 axis,
and it is graphed only for t &gt; 0. As can be seen from Figure 4b, g(t) can be
separated from the cross product terms in s(t) if T - T2 &gt; T2 - T1.  Since T2
is the point in g(t) nearest to the reference function, this condition implies
that the duration from T2 to the reference function must be greater than the
duration of g(t).  A similar analysis using Figures 4c and 4d when the reference function precedes g(t) again reveals that the duration from the closest
point in g(t) to the reference function must be greater than the duration of
g(t) in order that the various terms in s(t) may be separated.
14



</P>
<P><PB REF="00000022.tif" SEQ="00000022" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="864" N="00000022">
g(t)
Ti   T2            T
Figure 4a. Graph of a discontinuity delimited by 8(t-T1) and 5(t-T2).
The reference function, 8(t-T), follows the discontinuity.
s(t)
T2-T, T-T2 T-T,
Figure 4b.  Graph of s(t) as derived from the g(t) in Figure 4a.
5(t-T2+T1) is a cross product term and the discontinuity is
delimited by 5(t-T+T2) and 8(t-T+T1). Note that the time reversal
changes in the order of the two delimiters.



</P>
<P><PB REF="00000023.tif" SEQ="00000023" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="854" N="00000023">
g(t)
T           T       I T2
Figure 4c.  Graph of a discontinuity delimited by F(t-Tl) and F(t-T2).
The reference function, 8(t-T), precedes the discontinuity.
s(t)
T2-T T1-T T2-T
Figure 4d. Graph of s(t) as derived from the g(t) in Figure 4c.
5(t-T2+T1) is a cross product term and the discontinuity is delimited by b(t-T1+T) and 8(t-T2+T). Note that there is no time
reversal.



</P>
<P><PB REF="00000024.tif" SEQ="00000024" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="867" N="00000024">
Figure 16 in Section III shows a situation where g(t) is not separated
from the cross product terms. Figure 16a shows two large targets delimiting
some smaller targets and followed by the reference function. In this example
T - T2 &lt; T2 - T1. The deconvolved data shown in Section IV in Figure 16b reveal two false targets, a and b, that are the cross products between target c
and targets d and e.  Targets c, d, and e and everything in between are all
true indications.
The Fourier inversion step referred to above is a digital Fourier transform (DFT) implemented on a digital computer using the recently developed fast
Fourier transform (FFT) technique. Thus the DFT of a time sequence, f(nT), is
denoted by F(kQ) and
M- 1inkT
F(kn)  =   E  f(nT) e-                                               (18)
m=O
and
1 N-l        inkT
f(nT)  =          F(kD) e                                            (19)
k=O
where Q = 23~/NT and T is the uniform sampling interval in the time domain.
Gold and Rader (1969, page 162) snow that M = N, that is, the number of discrete frequency samples generateu by a DFT is equal to the number of discrete
time samples.  They also show that
DFT(e     )  =  Nq                                                   (20)
Thus, the DFT has the same frequency selective properties as the normal Fourier
transform. The factor of N appearing on the right side of equation (20) distinguishes the DFT from the conventional Fourier transform (FT) since
/ei 2t f t)
FT  e 2      )   =  b(2tf - 2if)                                     (21)
Thus all power terms which are derived from a function processed by a DFT will
be a factor of N2 greater than those processed by a conventional Fourier
transform.
SIGNAL-TO-NOISE RATIO ANALYSIS
Up to now, all the signals in the ultrasonic and data processing systems
have been assumed to be completely deterministic.  This assumption is now
17



</P>
<P><PB REF="00000025.tif" SEQ="00000025" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="865" N="00000025">
discarded in order to derive a formula expressing the SNR of the combined
ultrasonic and data processing system in terms of the SNR of a conventional
ultrasonic system.
There are two important noise sources in an ultrasonic system: reflections from the randomly distributed grain structure and thermal noise in the
ultrasonic transducer.  Since the target is usually much larger than the grain
structure of the medium, the random fluctuations in output in the vicinity of
a target are negligible.  Thus primarily, the random acoustic reflections
scatter the acoustic energy out of the main lobe of the transducer radiation
pattern and reduce the output voltage of the transducer. Therefore, the following analysis assumes that the thermal noise is the primary noise source.
As Davenport and Root (1958, page 185) state, thermal noise can be represented
by a, Gaussian process with a flat power spectrum.
The SdNR for a conventional ultrasonic system is considered first.  This
system is modeled in Figure 5a.  From the model it can be shown (Seydel, 1972,
page 47) that the signal-to-noise ratio for the unprocessed data is
1/2 Z A2
m m
U       2N B
o o
where N0 is the spectral density of the noise, B0 is the bandwidth of the
filter, P(f), and the numerator is signal power.
The model of the ultrasonic and data. processing system is shown in Figure
5b.  The wave analyzer is modeled as a tunable bandpass filter, K(f'), followed
by a square law device and a low pass filter, L(f).  After the bandpass filter
is tuned over the bandwidth of the ultrasonic signal, the resulting ensemble
of frequency samples is Fourier transformed to yield the final output.  The
deconvolution filter is omitted from this analysis as its effect will be considered below.
An analysis of the model (Seydel, 1972, page 47) shows that the SNR of
the processed data, (SNR)p, is
(SNR)2          B
(SNR)   =                    LBj2+B]                                 (22)
where B0 is the bandwidth of the ultrasonic transducer, B1 is the bandwidth of
the wave analyzer, IF filtr,, TB2 is the bandwidth of the low-pass filter, and
M is the number of samples taken from the power spectrum.
18



</P>
<P><PB REF="00000026.tif" SEQ="00000026" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="827" N="00000026">
nl(t)
Q(f)
q(t)
Figure 5a. Model of a conventional pulse-echo ultrasonic system with
output fl(t) combined with additive noise nl(t) and inserted into a
rectangular bandpass filter Q(f) designed to optimize the SNR.
n,(t)
K(f)                                       L(t)
f lt                BANDPASS                       2           LOW PASS  DIGITAL.I(B  FILTER                    ix4,FiTERASFOURI
x(t)                 x(t)                  x3()    FILTER                                x RANS(t)
K(t)                                       I (t)
Figure 5b. Model of the data processing system. The output from the
ultrasonic system, fl(t), is combined with additive noise, nl(t), to
form xl(t).



</P>
<P><PB REF="00000027.tif" SEQ="00000027" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="872" N="00000027">
Equation (23) is the formula relating the SNR of the data processing
system to the SNR for the conventional ultrasonic system. This formula can be
simplified slightly by considering some typical numbers for the various parameters.  In the experimental work presented below, the following values for M,
Bo, B1, and B2 are typical.
M  =  256      B   =  2.0 MHz       B1     200 Hz      B2    1 H    (24)
o                   1                   2        z
Then
B1        -6
MB   &lt;  10                                                           (25)
o
and this term will be considered negligible.  Thus
(SNR)          B           B
=  5ooo                          (26)
(SNR)U     2(B +B2)       2B1                                        (26)
and the data processing system introduces a 37 dB increase in the,SNR.
Note that both M and B2 appear to have very little effect on tqe final
SNR.  This is not true.  If B2 is not explicitly provided as a filter external
to the wave analyzer, it still has to be included as the low pass filter inside
the wave analyzer which normally follows the square law device. In a typical
high frequency wave analyzer, B2 is about 10 kHz, but without anottner noise
source, any B2 which is greater than 200 Hz only decreases the SNc by 3 dB.
In fact, the amplifiers in the wave analyzer introduce a noise source not included in the above analysis.  In this case, both M and B2 increase the SNR
more than shown in equation (23). The narrow bandwidth of B2 averages the
voltage over time, and the factor of M is due to the Fourier transform providing a weighted average over the ensemble of power samples.
The formal mathematics tends to obscure the physical basis of this increase in SNR.  It is merely due to the fact that the ultrasonic system is a
pulsed, evoked response system.  In such a system, the pulsed signal always
occurs at the same point in time while the noise at this point in time fluctuates in value for each succeeding pulse. If for a. given point in time, the
voltage produced by each pulse is added onto the voltage produced by the previous pulse, then the signal voltage is cumulative while the noise voltage
fluctuates and eventually cancels out. In the data processing system, the
narrow bandwidths of B1 and B2 provide this additive or averaging effect.
20



</P>
<P><PB REF="00000028.tif" SEQ="00000028" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="889" N="00000028">
INTERPRETATION OF THE ANALYSIS
As equation (15) demonstrates, the output of a data processing system consisting of a wave analyzer, an analog-to-digital converter, a digital deconvolution filter, and a Fourier transformer contains the reflectivity function,
g(t), in a form that can be easily recovered by a time-gating process.  An
output that is a direct measure of the reflectivity function has many advantages over a conventional ultrasonic system where the reflectivity function is
modified by the transducer.  These advantages, as well as some other responses
characteristic of this form of the output, are discussed below.
Perhaps the most obvious advantage of measuring the reflectivity directly
is that the format of the output is independent of the particular transducer
used in the pulse-echo system.  That is, a single point target is always displayed as a single spike at the output regardless of which transducer is u.sed
in the ultrasonic system. In a conventional system, a single point target is
displayed in a wide variety of forms. This statement concerning transducer
independence can be made even stronger if the transducer bandwidth is taken
into consideration. For a collection of transducers of approximately equal
bandwidth, not only is the output format the same, but the output function
representing a complex target is virtually identical for each transducer.
Transducer independent output is important when test results taken over a
period of years are compared for possible changes in flaw structure.  Over
this period of time, the piezoelectric crystal can age (Mason, 1950, page 100)
or be replaced due to failure.  Thus, in a conventional system, the flaw structure could appear to be modified when, in fact, the modification was strictly
due to a change in transducer response. The data processing system output is
not susceptible to this problem.
To a certain extent, this transducer independent output format is achieved
in a conventional pulse-echo system by the rectification and low-pass filtering
of the pulse before it is displayed.  But this process suffers in comparison
with the present data processing system because it achieves transducer independence at the expense of decreased resolution.
Another advantage is the increased resolving power provided by the data
processing system. The impulse response of an ultrasonic transducer can be
nicely approximated by an exponentially damped sinusoidal wave. With this
assumption, plus the fact that accurate spectral measurements can be made at
frequencies where the power is 20 dB below the power at the transducer center
frequency, it is easy to show that the bandwidth of the deconvolved data is
ten times the bandwidth of the unprocessed data. Since the bandwidth is
directly proportional to the resolving power, the data processing system would
have ten times the resolving power of the conventional ultrasonic system. In
practice, the increase in resolving power for the processed data is usually
between a factor of four and a factor of six.
21



</P>
<P><PB REF="00000029.tif" SEQ="00000029" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="884" N="00000029">
Cox and Renken (1970) have considered the application of a deconvolution
technique to electromagnetic transducers.  They feel that the technique is not
practical because the deconvolution filter is nonrealizable, amplifies small
impulsive disturbances at the input, and reduces the SNR to zero. The deconvolution filter is nonrealizable only when implemented in the time domain.  If
the filter is constructed in the frequency domain as described above, no realizability problems are presented. Their second objection is really just a
restatement of the third. Since the deconvolution filter is preceded by a
transducer that smooths out any impulses applied to its input, the only possible source of such impulses is the transducer noise.  But as was shown in the
above section on system noise, the SNR at the deconvolution filter input is
very nigh, even for a low SNR at the transducer.  In order to calculate the
SNR loss duae to the deconvolution filter, it is necessary to know tihe explicit
functional form of the transducer response.  Thus it is not possible to make
any general statements concerning the SNR at the deconvolution filter output.
From the fact that experimental SNR improvement factors within 10 dB of tue
theoretical value have been achieved, it is believed that the SNR at the output of the deconvolution filter approximates the value predicted by equation
(22,) and represents a. significant increase over what is expected from a conventional system.
While the data processing system output has both an improved resolving
power and an improved SNR with respect to a conventional system, t-ie usual
trade-off between resolving power and SNR still exists. But now, the conflict
has been escalated to a. higher level.  Figure 17 in Section IV is an example
where errors in the deconvolution filter have decreased the SNR.  Figure 10 is
the same transducer located in a slightly different position with a properly
constructed deconvolution filter.  The resolution in Figure 17 is slightly
better, but the SINR is badly degraded.  This indicates that the resolution-SNR
trade-off for the data processing system may be more severe than for the conventional system.
Equation (15) also demonstrates that both the phase and the a.mplitude of
g(t) have been preserved.  This opens up the possibility of forming a synthetic
array by taking data from many different transducer positions, processing the
data into a form described by equation (15), and then summing the data with
the appropriate phase. This technique would provide an extremely flexible display method since the array could be focused and directed under computer control after all the data has been collected. Thus, the computer processed data
is in an extremely advantageous form for processing the data returned from the
scanning system proposed in Section I.
22



</P>
<P><PB REF="00000030.tif" SEQ="00000030" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="883" N="00000030">
III.  EXPERIMENTAL PROCEDURE
EQUIPMENT
A schematic diagram of the complete ultrasonic and data processing system
is shown in Figure 6.
From an experimental point of view, the most important element in Figure
6 is the l-MHz reference oscillator.  This oscillator links the wave analyzer,
the pulser, and the range gates into a servo loop, and this servo loop provides the stability that enables highly accurate, drift-free measurements to
be made on the ultrasonic data.
The reference oscillator provides a stable reference frequency (1 part in
106) for use within the three phase locked loops (PLL) contained within the
analyzer and also, through the divider and timing circuits, triggers the ultrasonic pulser. Since the ultrasonic signal can be written as a Fourier series,
then if the wave analyzer is tuned to the center frequency of one of the comb
spikes comprising the signal spectrum, the servo loop locks the wave analyzer
to this frequency. If the reference oscillator drifts, both the ultrasonic
pulse rate frequency and the wave analyzer center frequency change in synchronism.  The PLL corrects any drift within the wave analyzer's three local oscillators to an accuracy of ~10 Hz.
The timing and divider circuits are used to trigger both the ultrasonic
pulser and the gates. The divider circuit consists of a divide by five and
two divides by ten digital counters that reduce the reference oscillator frequency to 2 kHz.  This 2-kHz signal is sent to the timing circuits where it
simultaneously initiates a trigger pulse for the ultrasonic pulser and triggers
the first of two adjustable monostable multivibrators arranged in series.
The first multivibrator sets the gate delay, and then it triggers the second
multivibrator which sets the gate width. The gate signal from the timing circuits is sent to a pair of double balanced mixers which perform the actual
gating operation.  When the gate is "on," the mixers pass the ultrasonic echoes
with only slight attenuation.  When the gate is "off," the mixers attenuate
the echoes (up to about 80 dB). In this manner, the gate passes only those
ultrasonic echoes of interest. The amplifier is inserted between the two
mixers in order to avoid overloading the first mixer on strong signals.  The
output from the second mixer is supplied to the wave analyzer and to the CRT,
where it is monitored.
The wave analyzer has three local oscillators and a square law detector
that provide an output voltage proportional to the power at its center frequency.  A local oscillator and mixer operating from 30 to 48 MHz tune the
instrument from 10 kHz to 18 MHz.  The constant difference frequency of 30 MiHz
25



</P>
<P><PB REF="00000031.tif" SEQ="00000031" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="742" N="00000031">
reference I
PIEZOELECTRIC
TRANSDUCER -
WAVE        LOW-PASS
GATE     AMPLIFIER       GATE
ULTRASONIC                                                         ANALYZER
PULSE._
4z I -        |TEST SPECIMEN
DISCONTINUITY
GATE     AMPLIFIE        GATE             CRT
VOLTMETER
INVE.RSE FOURIER               DIGITAL,COMPUTER
TRANSFORM             DECONVOLUTION               LINKAGE
Figure 6. Schematic diagrami of the complete ultrasonic and data processing systems.



</P>
<P><PB REF="00000032.tif" SEQ="00000032" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="884" N="00000032">
is then mixed with another 30-MHz local oscillator.  The difference frequency
is filtered by a low-pass filter (typically 200-Hz bandwidth), mixed with a
250-kHz local oscillator, and applied to a square law detector. The output
from the square law detector is proportional to the power in a 200-Hz bandwidth centered on the chosen frequency. The frequency tuning is indicated by
a seven-digit counter.
The wave analyzer detector contains a 10-kHz bandwidth, low-pass filter.
This filter has too much bandwidth to adequately limit the amplifier noise in
the analyzer, so the detector output is additionally filtered by a 1-Hz bandwidth, third order Butterworth realization of an active, low-pass filter. An
integrating digital voltmeter (DVM) capable of 10-bit resolution, digitizes
the output from the 1-Hz filter and this digital data is entered directly into
a storage file on a large, central computer using a remote terminal with an
acoustic coupler.
The accuracy specification for the DVM was chosen by taking readings at
four test frequencies over a period of about four hours.  It was found that
for 8-bit accuracy. the drift was about +1/2 least significant bit.  Also, as
Gold and Rader (1969, page 101) show, the quantization noise variance for
8-bit accuracy is about 60 dB below the peak signal. This SNR is about equal
to the SNR from the transducer thermal noise in a minimal noise situation.
Thus, 8-bit accuracy is necessary to maintain the assumption that transducer
thermal noise is the primary noise source.
For all the data that are reported here, the reference function is provided by the back surface reflection. A duplicate gating and amplifier system adjusts the reference function amplitude independently of the target amplitude. Two more monostable multivibrators together with another pair of
double balanced mixers gate the back surface reflection to the summing junction.  The amplifier gain is adjusted to make the reference function amplitude
slightly greater than the maximum target echo amplitude. The target signal
and the reference signal are additively combined and sent to the wave analyzer
and to the CRT monitor.
A FORTRAN computer program controls the data conversion, deconvolution,
and inverse Fourier transform procedures. Since the wave analyzer measures
the power in discrete 10-dB bands, the DVM data contains a scale factor that
must be used to calculate the actual power. This conversion process could
also compensate for any nonlinearities in the wave analyzer, but it was experimentally established that the effect of such nonlinearities was less than
the quantization noise and no compensation was used. The linearity was measured by inserting a stable (in amplitude and frequency) l-MHz sine wave through
a precision attenuator box into the wave analyzer.  The linearity was measured
only at this one frequency.
25



</P>
<P><PB REF="00000033.tif" SEQ="00000033" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="880" N="00000033">
After conversion, the data are deconvolved using point by point division.
This result is multiplied by a Hamming window function and Fourier transformed
back into the time domain using an FFT algorithm.  The final results are available as a sequence of numbers relating the absolute value of the amplitude and
time or as a Calcomp graph of absolute amplitude versus time.  The graph omits
the data near t = 0 where the cross product terms are located.
At the present time, the program is interactive, as information about.initial,  final, and incremental frequency inmust oe supplied.  An interactive
program was chosen for its experimental flexibility, but an automatic system
seems just as feasible.
TEST SAMPLES
Two basic test targets established the feasibility of this system.  The
target in Figure 7a consists of three steps, B, C, and D, milled into a 150-mm
high aluminum block.  The steps are 3. 13 mm wide and separated by 0. 508 mm and
0.787 mm, respectively.  These steps are intended to simulate closely spaced
stringer inclusions.  Step A or the back surface provide a reference function,
depending on the transducer location. The target in Figure 7b consists of
three, 1.6 mm, side-drilled holes, E, F, and G, randomly arranged in a 63.5-mr
high aluminum block.  The back surface provides the reference function.  This
target system is intended to simulate porosity. Both test targets are in the
far field of all the transducers used in the following experiments.
An additional test target (not shown) is used to establish the expected
SOiR improvement.  It consists of a 13-mm diameter, flat-bottom hole drilled
into a plastic block loaded with metallic particles. The metallic particles
scatter considerable energy out of the main beam, and, as a result, the echo
from the flat-bottom hole is extremely weak and submerged in the transducer
thermal noise.
The defect distance is defined as one half the acoustical path from the
transducer to the reference target and back again minus one half the acoustical
path from the transducer to the defect target and back again. Thus, defect
distance equals the distance between the defect target and the reference target slightly modified by an obliquity factor.  Defect distance is measured
using both an ultrasonic and geometrical technique. The formula d = 1/2 cT
is used to calculate the ultrasonic defect distance, where c is the ultrasonic velocity of propagation and T is the time of flight to the reference
target minus the time of flight to the defect target.  The geometrical defect
distance is calculated using data provided by micrometer measurements of target position. For the three-step target in Figure 7a, a depth micrometer is
used to measure the distance from each step to the back surface, and a standard micrometer is used to measure the total height of the block. For the
three-hole target in Figure 7b, an optical microscope equipped with a micrometer
translation stage is used to measure target position.
26



</P>
<P><PB REF="00000034.tif" SEQ="00000034" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="881" N="00000034">
E
oFG
Figure 7a.  Targets B, C, and D          Figure 7b.  Targets E, F, and G
consist of a 0.508-mm and a              consist of three 1.6-mm diameter
0.787-mm step in aluminum.              holes drilled in aluminum. Back
Steps A and E supply a refer-           surface supplies the reference
ence function.                          function.
27



</P>
<P><PB REF="00000035.tif" SEQ="00000035" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="884" N="00000035">
For the three-step target in Figure (a, four different transducer positions are used. In the first position, denoted as position 1, the transducer
center line is coincident with the left side of step B.  In the second position, denoted as position 2, the transducer center line is coinci i &mdash;nt with
the right side of step B.  In the third position, denoted as position 3, the
transducer center line is coincident with the right side of step C. And in
the fourth position, denoted as position   -, the transducer center line is
coincident with the right side of step D.
Since all the targets are located in the far field of the transducer,
they are all insonified with a plane foa.re, but not all the targets will reflect a plane wave.  The type of reflection process occurring when the three
steps are insonified can be seen by considering an infinite, smooth surface
with a step at the origin.  For a transducer located far to the left of the
origin, the surface is insonified with a plane wave, and, since there are no
irregularities within the beamwidth of the transducer, a plane wave is reflected back to the transducer.  This situation remains the same as the transducer is scanned toward the origin until the step comes within the transducer's
beamwidth.  Then both the upper and lower surfaces reflect the incident pulse,
but the pulse from the upper surface is not received due to the finite size of
the transducer. At this point, the upper surface would not be detected were
it not for the fact that the corner of the step reflects a spherical wave.
This spherical wave is received by the transducer (with an amplitude diminished
by the inverse distance law), and the presence of the upper surface is detected,
but with a diminished amplitude relative to the lower surface.  As the scan
is continued to the right, the plane wave pulse reflected from the upper surface is finally received by the transducer and both surfaces are indicated by
about the same amplitude.
This analysis is important in calculating the acoustic distance to each
step of' the three-step target.  If both surfaces of a step are reflecting
plane waves, then the acoustic distance is the perpendicular distance from
the transducer to the surface, plus the perpendicular distance from the surface
to the transducer.  But if the upper surface of the step is detected by the
spherical wave reflected from the corner, then the acoustic distance is the
perpendicular distance from the transducer to the step plus the oblique distance from the step to the transducer.  Thus, for position 1 for the threestep target, the transducer is located directly over step A (the reference
function) and. step B, and these steps reflect plane waves. The plane waves
reflected by steps C and D miss the transducer and these steps are detected
only by the spherical waves reflected by the corner of each step. Likewise,
in position 2, steps B and C reflect plane waves and step D and Step A (the
reference function) reflect spherical waves; in position 3, steps C and D
reflect plane waves and step B and step E (the reference function) reflect
spherical waves; and in position 4, step E (the reference function) and step
D reflect plane waves, and steps B and C reflect spherical waves.  As is shown
later in Figure 13, a small amplitude spherical wave reflection can be seen
28



</P>
<P><PB REF="00000036.tif" SEQ="00000036" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="887" N="00000036">
from step A with the transducer in position 4. These transducer locations
are noted in Table I and in the captions for Figures 9-13, and they account
for the slightly different distances assigned to the same target for different
transducer locations.
Uncertainties in the location of the peak value of the target display
introduce an error in the ultrasonic distance measurement of approximately
~0.05 mm.  Uncertainties in the transducer location introduce an error in the
geometrical distance measurement of approximately +O. 075 mm. The micrometers
used to provide the data for the geometrical distance calculations are all
calibrated on a regular basis, and their error is assumed to be less than
~12 [Em. For this reason, micrometer error will be neglected.
The ultrasonic velocity was independently measured using the test block
shown in Figure 7a.  Since the test block is only 26. 1 mm thick, multiple reflections of one pulse within the test piece continue for four or five round
trips when the transducer is placed on the front face. If two of these multiple reflections are gated into a spectrum analyzer, a sine wave modulated
spectrum results. If as many multiple reflections as possible are gated into
the spectrum analyzer, the sine wave modulation is "sharpened" into a series
of spikes in the same way that a Fabry-Perot etalon "sharpens" the frequency
selectivity of an optical interferometer.
A spectrum analyzer recording of this process is shown in Figure 8. The
upper trace is the "sharpened" spectrum, the lower trace is a calibrated frequency comb. The frequency difference, A, between adjacent spikes is related
to the ultrasonic velocity, c, and the thickness, d, according to the following
formula.
c  =  2dA
A was measured by averaging over 22 different spikes and d was measured using
a micrometer.  A, and thus c, is estimated to be accurate to within ~0.5%.
The value of c derived from this formula and used in all subsequent calculations is 6.28 x 106 mm/second.
PROCEDURE
The system shown in Figure 6 is operated in the following manner.
1.  Place the transducer over a single point target.  Take the echo return and apply it to the wave analyzer. Tune the wave analyzer to
each comb tooth in the echo spectrum, digitize the measured power,
and store in a computer for subsequent use.
29



</P>
<P><PB REF="00000037.tif" SEQ="00000037" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="821" N="00000037">
TABLE I
COMPARISON BETWEEN ULTRASONIC AND MICROMETER DISTANCE MEASUREMENTS
Ultrasonic        Geometr i cal
Figure    Position    Target
Distance in mm    Distance in mm
9          2           B             7.90               7.75
C            7. 06              6. 96
D            6.43               6. 38
10          2           B            7.98                7. 75
C            7.21               6. 96
D            6.58               6.38
11          1           B            7. 72               7.67
C            6. 99              6.81
D            6.25               6.22
12          3           B           14, 1               14. o
D           12. 7              12.8
13          4           B           13. 7               13.8
D           12.6               12. 7
14                      El  2 10.2                      10. 2
F            8.59               8.59
G            7.47               7.16
15         *            E           10. 1               10. 2
F            8.59               8.59
G            7.21               7.16
17          4           B           13.8                13.9
C           13. 2              13. 2
D           12.6               12. 7
30



</P>
<P><PB REF="00000038.tif" SEQ="00000038" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="784" N="00000038">
1!
-  111- - IAA      A    I A  Il    I -   A             I I
9 MHz                    10 MHz                     11 MHz
Figure 8. Power spectrum of the multiple reflections in a 26.1-mm thick
aluminum block.  Small pips on lower trace occur every 100 kHz.



</P>
<P><PB REF="00000039.tif" SEQ="00000039" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="886" N="00000039">
2.  Place the transducer over the unknown target.  Using one gate system
for the target echo and one gate system for the reference function,
add these two signals and apply the combination to the wave analyzer.
Tune the wave analyzer over the spectrum, digitize, and store the
result in a computer as in step 1.
3.  Point by point divide the iigital power spectrum from step 2 by the
digital power spectrum from step 1.
4.  Multiply the result from step 3 by a Hamming window function and
Fourier transform the signal back into the time domain.  Now the
signal can be displayed in any of the conventional ways (CRT, strip
chart recorder, etc.).
Step 1 measures the function Sl(f) in equation (8). If the medium has a
negligible effect, Sl(f) is the transducer power spectrum, sometimes called
the transducer frequency response. Step 2 measures the function S2(f) in
equation (12). Step 3 is the deconvolution procedure. In step 4, the Hamming
window function weights the spectral data so as to provide minimal side-lobe
level in the final time-domain display. Figure 12 shows an example where the
side-lobe level is 32 dB below the peak signal, as compared with a predicted
value of 42 dB.
32



</P>
<P><PB REF="00000040.tif" SEQ="00000040" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="887" N="00000040">
IV. RESULTS
Figures 9-18 display the results obtained on the two test targets in
Figure 7 and on the flat-bottom hole in the plastic block.  Figures 9 through
13 use the three-step target in Figure 7a, Figures 14 and 15 use the threehole target in Figure 7b, and Figure 18 uses the flat-bottom hole in the
plastic block,  Figures 16 and 17 are included primarily to illustrate points
made in the theoretical development of the method.  Four different transducers,
two 10-YlHz units, a 5-MHz unit, and a 2.25-Ivlz unit, were used in this
experimental study. Their responses are shown in Figure 2. Both 10-MHz
units are 6.4 mm in diameter.  One unit, designated SFZ, is a zirconatetitanate ceramic unit with an impulse response shown in Figure 2b.  The other
unit, designated SCJ, is a special faced-ceramic unit with an impulse response
pictured in Figure 2d.  The exact construction of the SCJ unit is proprietary
and hence no information about the unit is available.  The 5-MHz unit is a
13-rmm diameter, lead metaniobate crystal with an impulse response pictured in
Figure 2c. The last transducer, a 13-mm diameter, 2.25-MHz zirconate-titanate
ceramic unit designated SFZ, has the impulse response shown in Figure 2a. The
particular transducer used in each instance is indicated in the figure caption.
With one exception, the format of Figures 9-18 is the same. The photographs in the figures show the CRT monitor (see Figure 6) taken during the
data acquisition phase of the experiment.  This waveform supplies the actual
raw data used in the computer processed data graphed beneath it. The computer
processed data are presented as a 65% reduction of a Calcomp graph of the
final, processed output. This output is also available in tabular form (not
included), and it is from this table that the distance data for each target
indication is calculated.  Each target indication on the Calcomp graph is
identified with a step or hole as labeled in Figure 7.  Since the reference
function is preceded by the simulated defect, the time scale on the computer
processed data is inverted.  On the oscilloscope photographs, the targets
appear in natural order the simulated defect precedes the reference function.
On the Calcomp graph, the reference function appears at t = 0 (it has been
masked off in order to emphasize the simulated defect), and the simulated
defect follows.  Thus, when temporal relationships between the two graphs are
being compared, either the Calcomp graph or the oscilloscope photograph must
be reflected through the reference function. With a little more sophisticated
manipulation of the Calcomp plotter, this problem could be avoided.  The time
scale of each graph is noted in the caption. The ordinate of the oscilloscope
photograph is relative amplitude.  Since the computer output is a complexvalued function, the Calcomp ordinate is the absolute value of the amplitude
normalized to the largest target indication. Detailed information on each
experiment is reported elsewhere (Seydel, 1972).
33



</P>
<P><PB REF="00000041.tif" SEQ="00000041" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="699" N="00000041">
TH-REE STEPS      REFEREC
Figure 9a.  RF "?A" scan presentainoth
three-~step, and  reference echoes  sn   h
10-MHz SFZ transducer in positio 2.Tm
B                        scale is 500 ns/div.
S                           ~~~~~~~~~C
D.00      so0    1.00    1.50o    2.00    2.50    3'.00     3 CRECND                                       I   4   450 S:0  5S
Figure 9b.  Computer drawn graph of the three-stepD target obtained by processing the data show   nFgr   a



</P>
<P><PB REF="00000042.tif" SEQ="00000042" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="678" N="00000042">
THREE STEPS        REFEEC
Figure l0a.  RF "A  scan presenainoth
three-step and  reference echoe   sn   h
1-0-MHz SCJ transducer in positio.Tm
scale is 500 nsjdiv.
S                             ~~~~~~~~~B
D
S                          ~~~~~~~~C.00.50     1.00     so5   2.00O    2.0 so   5.00    5.50    4.00    4.0     S.00    S.50o    6. 00.0     70
MICROSECONDS
Figure l0b.  Computer drawn graph, of the three-step target obtained by processing the data show   nFgueia



</P>
<P><PB REF="00000043.tif" SEQ="00000043" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="701" N="00000043">
THREE STEPS         REFERENCE
Figure Ia.   RF "A" scan presentation of the
three-step and reference echoes using the
\IN                                                                         5-MHz lead metaniobate transducer in position
B~~~~~
8  B~~~~~~~~~~~~~~1  Time scale is 500 ns/div.
C
D
8 i.00     1.00    2.00    3.00    4.00    S.00    6.00?.CO    6.00    9.00    10.00    11.00    12.00    13.00   1u4.00
MICROSECOONDS
Figure llb.  Computer drawn graph of the three-step target obtained by processing the data shown in Figure ha.



</P>
<P><PB REF="00000044.tif" SEQ="00000044" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="737" N="00000044">
THREE STEPS         REFERENCE
Figure 12a.  RF "A" scan presentation of the
three-step and reference echoes using the
10-MHz SCJ transducer in position 5. Time
D                                  scale is 1.0 p~sec/div.
S"                                D
AP                         ~~~~~~~B.00     1.00    2.00    3.00    4.00    S.00    6.00    7. 00    B.00    9.00    10.00    11.00    12.00    13.00    14.00
MICROSECONOS
Figure 12b.  Computer drawn graph of the three-step target obtained by processing the data shown in Figure 12a.



</P>
<P><PB REF="00000045.tif" SEQ="00000045" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="794" N="00000045">
THREE STEPS         REFERENCE
Figure 13a. RF "A" scan presentation of the
s       8~D                                                                    three-step  and reference echoes using a 5MHz lead metaniobate transducer in position
4.  Time scale is 1.0 isec/div..2.
8f.00     1.00    2.00    3.00    4.00    S.o00    6.00    7.00    8.00    9.00    10.00    11.00    12.00    13.00    14.
HI CROSECONDS
Figure 13b. Computer drawn graph of the three-step target obtained by processing the data shown in Figure 13a.



</P>
<P><PB REF="00000046.tif" SEQ="00000046" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="746" N="00000046">
THREE HOLES         REFERENCE
Figure 14a. RF "A" scan presentation of the
three-hole  and  reference echoes using a 5F                                              MHz lead metaniobate transducer.  Time scale
is 500 ns/div.
8                           E
cr8                  G
mult iple
ref lection.00     1.00    2.00     3.00    q. 00    S. 00    g,.0     1.00   a 00    9.00    10'.00    11.00    12.0I1o          uo
MICROSECONOS
Figure 14b.  Computer drawn graph of the three-step target obtained by processing the data  hw      Fgr  La



</P>
<P><PB REF="00000047.tif" SEQ="00000047" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="802" N="00000047">
THREE HOLES        REFERENCE
Figure l5a. RF "A" scan presentation of the
three-hole and reference echoes using the
8S                     F                                           2.25-MHz SFZ transducer.  Time scale is 1.0
o                                                                          4lsec/div.
E
8.00     1.00    2.00    3.00    4.00    5.00    6.00    7.)0   68.00    9.00    10.00   11.00   12.00   13.00    14.00
MICROSECONDS
Figure 15b.  Computer drawn graph of the three-step target obtained by processing the data shown in Figure 15a.



</P>
<P><PB REF="00000048.tif" SEQ="00000048" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="783" N="00000048">
Figure 16a. RF "A" scan presentation of the
target and reference echoes from a cast iron
sample using the 2.25-MHz SFZ transducer.
C
8-                                          C                                    Time scale is 2.0  Lsec/div.
a
e
b
8'                    4.00     2.00'.00     6.00     8.00     t0.00    12.00    1.00     6.00    te.00    20.00    200    24.. 00
MICFOSECONDS
Figure 16b.  Computer drawn graph of cast iron target obtained by processing the data shown in Figure 16a.



</P>
<P><PB REF="00000049.tif" SEQ="00000049" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="739" N="00000049">
THREE STEPS        REFERENCE
Figure 17a. RF "A" scan presentation of tne
three-step and reference echoes using the
10-IHz SCJ transducer.  Time scale is 1.0
iD Lsec/div.
S~~~~~~~~~~~~~ C
o                                                                         B.00.50      1.00    1.50    2.00    2.50    3.00      3.S)'.0O     4.5O    5.00    5.50    6.00    6.50    7.00
MICROSECONDS
Figure 17b. Computer drawn graph of the three-step target obtained by processing the data shown in Figure 17a.



</P>
<P><PB REF="00000050.tif" SEQ="00000050" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="788" N="00000050">
MILLED STEP         REFERENCE
Figure 18a.  RF "A" scan presentation of the
target and reference echoes from the plastic
block using a 5-MHz lead metaniobate transducer.  Time scale is 2.0 Cisec/div..00     2.00    4.00    8.00    8.00    I0.00   12.00    t4.)    16.00.00   20.00    22.00    24.00    26.00    2.00
MI CROSECNO$DS
Figure l8b.   Conputer drawn graph of the milled step in the plastic block obtained from the data in Figure  8a.



</P>
<P><PB REF="00000051.tif" SEQ="00000051" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="879" N="00000051">
V. DISCUSSION OF RESULTS
The computer processed data displayed in Figures 9-18 is superior to a
conventional "A' scan display in four crucial areas, namely, the accurate
measurement of target distance, transducer longitudinal resolution, transducer
independence, and transducer SNR.  These are considered in detail below.
D IS TANCE MEASUREMENT
Table I (p. 30) displayed the ability of the data processing system to provide accurate measurements of the distance to a defect. As discussed in Section
III, the ultrasonic distance measurement has an error bound of ~0.05 mm, and
the geometrical distance measurement has an error bound of ~0.o8 mm. An
uncertainty of +0.5%0 in the ultrasonic velocity also contributes to the error
in the ultrasonic distance. Since the velocity error is the same for all
common paths, this error applies only to the distance between the reference
and defect targets. For defect targets whose distance is about 7.0 mm the
ultrasonic velocity error bound is ~0.08 mm.  Thus, the error bound of the
ultrasonic distance is  -0.13 mm, and the error bound of the geometrical
distance is +0.08 mm.  Of the twenty-two entries in Table I, nineteen of them
are within the combined error bound of ~0.21 mm.  If the error bound is lowered
to ~0.15 mm, seventeen of the twenty-two entries are included, and if the
error bound is lowered to ~0.10 mm, sixteen of the twenty-two entries are
included.  A majority of the excluded entries comes from Figure 10.  For this
example, the average relative error is only 0.025 mm while the average absolute
error is 0.23 mm. It could be contended that this large absolute error is
due to a thick wear plate on this particular transducer, since the wear plate
thickness is measured by the ultrasonic system but not by the micrometer,
but the same transducer is used in Figures 12 and 17, and neither of these
samples shows the same effect.
In a conventional ultrasonic system, comparable measurements could be
made by choosing a particular timing point (such as an axis crossing) on the
waveform of the returning echo.  In a medium with little frequency dependent
attenuation, the waveform will retain the same shape, and such a system can
equal or exceed the accuracy achieved by the data presented here. But when
comparing the two systems, three points must be kept in mind. First, for
comparison purposes, the appropriate error bound on the processed ultrasonic
data is ~0.05 mm since both systems are subject to the same error in ultrasonic
velocity. Second, because of the error bounds on the other quantities, no
attempt was made to increase the accuracy of the processed data by using
numerical interpolation.  And third, any frequency-dependent attenuation in
the medium causes a distance-dependent error in the conventional system.
Serabian (1968) discusses this problem and presents some experimental data to
44



</P>
<P><PB REF="00000052.tif" SEQ="00000052" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="875" N="00000052">
verify the effect.  But due to the media compensating qualities of the processing system, the processed data has an error which is a function only of
the relative distance between the reference and defect targets and not of the
absolute distance to the defect target.
RESO LUTION
As shown in Section II, the transducer longitudinal resolution can be
improved by as much as a factor of ten using a deconvolution procedure.  The
best examples of such an improvement in resolving power are Figures 9, 11,
and 15.  In Figure 9a, the back surface reflection shows that the transducer
rings for about 0.6 Nsec.  This time duration implies a distance resolution
of about 2.0 mm, a reasonable figure since the conventional "A" scan display
in Figure 9a does not resolve the three-step targets with a separation of
1.37 rm. The computer processed display in Figure 9b completely resolves all
three steps, and, thus, it demonstrates a resolution greater than 0.58 mm
(from Table I).  Using the fact that the maximum resolution equals the velocity
of propagation divided by twice the bandwidth (c/2BW), the computer processed
data in Figure 9b has a maximum resolution of 0.36 mm.  In this example, the
deconvolution process increases the resolution by a factor of six. In Figure
lla, the back surface reflection shows a resolution of about 0.6 usec or
about 2.0 mm. The computer processed data barely resolves all three steps,
and it thus demonstrates a resolution greater than 0.58 mm. In fact, the
resolution is probably more nearly 0.50 mm since the 0.58-mm step has about
one quarter the amplitude of the largest echo and resolution is usually
defined for equal strength targets. Using this assumption, the deconvolution
process increases the resolution by a factor of four. In Figure 15a, the
back surface reflection shows a resolution of about 1.5 usec or about 5.1 mm.
The computer processed data completely resolves all three holes, and it thus
demonstrates a resolution greater than 1.4 mm. Again, if the expression
c,/2BW is used to establish the greatest possible resolution, the computer
-processed data in Figure 15b has a maximum resolution of about 1.0 mm.  In
this example, the deconvolution process increases the resolution by a factor
of five.
Figures 10, 12, 13, and 14 also demonstrate corresponding increases in
resolution. In Figures 12 and 13, all three steps are not resolved even
though, as the data in Table II show, the maximum resolution is adequate.
It is believed that for these two cases, the transducer was placed in such a
position that the echo returned from one or two of the steps was extremely
weak in relation to the primary echo.
45



</P>
<P><PB REF="00000053.tif" SEQ="00000053" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="888" N="00000053">
TABLE II
MAXIMUM RESOLUTION EXPECTED FOR THE COMPUTER PROCESSED
DATA IN FIGURES 9-15
Figure                c/2BW in mm
9                     0.36
10                     0.41
11                     o. 48
12                     0.38
13                     0.41
14                     0.99
15                     0.97
Thus, the computer processed data have about five times the resolution
of the corresponding unprocessed data.  This falls short of the factor of ten
improvement expected from the deconvolution procedure because the power
spectral response of the transducers used in this experiment decreases more
rapidly than the Lorentzian response postulated in the theoretical analysis.
The resolution could be further increased by extending the bandwidth over
which the deconvolution procedure is applied. For the experimental data
presented here, the bandwidth was determined by the maximum and minimum power
levels accepted by the wave analyzer.  The upper power level was determined
by the necessity of limiting the intermodulation products, and the lower
power level was determined by the instrument sensitivity. For a particular
transducer, these two power levels determined the upper and lower frequency
cutoffs and hence the deconvolution bandwidth. If the deconvolution products
could be decreased or the sensitivity increased, the resolution could be
correspondingly increased.
The above conclusions concerning resolution improvement are drawn by
comparing the RF "A" scan display with the computer processed data. If the
comparison is made between the processed data and a rectified, filtered
display, the resolution improvement is even more dramatic. Figure 19 shows
a rectified, filtered display of the three-hole target using the same 5 MHz
transducer used in Figure 14. The transducer is located directly above the
center hole, exactly as in Figure 14. While the rectified display shows the
possibility of two different targets, it does not meet the industry specifications for resolution.  Thus, the rectified video display does not resolve
two targets 3.1 mm apart while the processed display has a maximum resolution
of 0.48 mm (From Table II, Figure 14).  This represents an improvement factor
of at least six, and, quite possibly, even as much as eight or ten.
46



</P>
<P><PB REF="00000054.tif" SEQ="00000054" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="878" N="00000054">
Figure 19.  Rectified video display of the
three-hole target in Figure 7b using the
5-THz transducer in Figure 2c.
TRANSDUCER- INDEPENDENT DISPLAY
The ability of the data processing system to provide an output display
that is almost independent of the ultrasonic transducer is seen by comparing
Figures 14 and 15 and Figures 10 and 11.  First, compare the target echo in
Figure 14a with the target echo in Figure 15a.  The echoes are dramatically
different mainly because the twco echoes come from a 5-MHz and a 2.25-MHz
transducer, respectively.  Also, the 5-MHz transducer almost resolves the
three holes while the 2.25-MHz transducer does not.  Next, compare the computer
processed data in Figure 14b with the computer processed data in Figure 15b.
Except for the strong multiple reflection in Figure 14b, the two displays are
qualitatively identical.  There are some quantitative differences in echo
strength and position probably due to minor differences in transducer location,
but the form of the display is the same for two transducers of different
operating frequency.  The same comparison process can be used for Figure 10,
a 10-MHz transducer, and Figure 11, a 5-MHz transducer, although in this case,
the quantitative differences are more substantial.
To a greater or lesser extent, all the computer processed data in Figures
9b-15b illustrate the same transducer independent display.  In every case, a
single target is represented by a single blip. The ringing response of the
transducer is almost completely removed since, in one example, the experimentally measured side-lobe level is 32 dB below the primary response.  Also,
this improved display technique is accompained with a simultaneous increase
in resolution. This distinguishes the present data processing technique from
the conventional rectified video display where display uniformity is achieved
only with a loss of resolution.
SN\R IMPROVEMENT
Figure 18 demonstrates the capability of the data processing system to
increase the SNR. Figure 18a is an "A'" scan presentation of the echo from
47



</P>
<P><PB REF="00000055.tif" SEQ="00000055" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="883" N="00000055">
a flat-bottom hole located in a plastic block loaded with metallic particles.
The metallic particles, having a much greater acoustic impedance than the
plastic, cause considerable scattering. This scattering decreases the echo
amplitude to the point where it is comparable to the thermal noise generated
in the transducer and associated electronics.  This condition is shown in
Figure 18a. Figure 18b shows the computer processed data. The data processing
used in this example is exactly the same as that used in all the previous
examples. The unprocessed signal is estimated to have a SNR of one. The SNR
for the processed data is difficult to estimate since certain sections of the
graph appear to be attributable to spurious signals. For instance, the blip
at 15 usec is almost certainly due to a slight imbalance in the balanced mixer
which allows the gate signal to transfer to the output, while the signal
between 4 and 8 4sec could be due to small imperfections on the flat-bottom
hole. If these ambiguities are included in the calculation of the noise
power, the SNR (peak signal to RMS noise) is 20 dB.  If the ambiguities are
excluded, the SNR is 28 dB.  The latter figure is probably more representative
of the system capability.  Thus, the data processing system has improved the
SNR by 28 dB, as compared with a predicted improvement of 37 dB.  A close
examination of the unprocessed data in Figure 18a reveals a pulse duration of
about 3.2 Nsec.  The corresponding time resolution in the processed display
is about 1.6 usec. Thus, the computer processed display shows a factor of two
improvement in resolution achieved simultaneously with a 28-dB improvement in
SNR.
The processed data in Figures 12 and 13 also demonstrated good SNR. In
Figure 12, the signal between the targets and the back surface reflection has
been gated out. Thus, the graph of the processed data from 1.0 to 3.5 Asec
gives a good indication of the amount of noise introduced by the deconvolution
process alone.  The SNR (peak signal to RMS noise) in this area is 37 dB.  In
Figure 13, the signal between the targets and the back surface reflection is
allowed to pass. The echo indications near 2.0 usec can be identified with
step A in Figure 7a.  In both figures, the SNR in the area following the
targets is excellent (about 45 dB).
COMPENSATION FOR A SCATTERING MEDIUM
The effect of a scattering medium upon the test results was briefly
considered in Section II. Equation (13) was derived to show the quantitative
effect of scattering on the data collected from the wave analyzer. Carrying
the derivation a few more steps shows that
S(f)  =  M2(f)M2 (f)G(f)G (f) +       1  +  e-i2ofT
M2 (f)(fG   ) + e    M2(f)G(f)                            (27)
48



</P>
<P><PB REF="00000056.tif" SEQ="00000056" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="863" N="00000056">
where S(f) is defined as before and
M  (f)
M2(f)       2                                                        (28)
M (f)
Equation (27) demonstrates the effect of a scattering medium on the
output from the complete data processing system. This effect will be illustrated by expressing equation (28) in terms of the material properties of
the aluminum test sample and inserting this result into equation (27) under
the condition that a single point discontinuity is insonified.
M2(f) can be expressed as a function of the material properties of
aluminum by using the equation relating attenuation and frequency shown below
(Mason, 1950, page 414),
~(f)  1=  B  f + B2f                                                 (29)
B1 and B2 are functions of the material structure.  Since a(f) is the attenuation coefficient expressed in dB/mm, then
(c(f)  =  [d log M(f)]                                              {(30)
where d is the distance of travel and M(f) is the usual amplitude transmission
function for a scattering medium as shown in Figure 1. Rearranging equation
(30) gives
-da(f)
M(f)     e                                                           (31)
and if equation (31) is substituted into equation (28), then
M2(f)  =  e-2(d1-d)c(f)                                              (32)
where d  - d is the distance from the discontinuity to the reference function.
In spite of the particular form of a(f), this model of the scattering
process demonstrates that the data processing system compensates for any
scattering which occurs over an acoustic path common to both the discontinuity
49



</P>
<P><PB REF="00000057.tif" SEQ="00000057" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="721" N="00000057">
al
8.00.50       1.00      1.50     2.00      2.50     3.00      3.50'.00'1so      S.00      5.50     6.00      6.50     7.00
MICROSECONDS
Figure 20.  Computer simulation of the effect of scattering on a single target with                              - d    0.0 mm.



</P>
<P><PB REF="00000058.tif" SEQ="00000058" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="692" N="00000058">
S
So..00.50        1.00      1. so     2.00      2:50      3.00      3.56      UO 4,0   1:SO      5.00      5. SO    ~0,0       t~
MICROSECONOS
Figure 21.  Computer simulation of the effect of scattering on a single target with di                                          1.    m



</P>
<P><PB REF="00000059.tif" SEQ="00000059" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="580" N="00000059">
8
S
S
8
0S
S
8.00.50       1.00      1.50     2.00      2.50     3.00      3.50      4.00     4.SO      5.00     S.SO      6.00      6.50     7.00
MICROSECONOS
Figure.    Comuter  simulation of the effect of scattering on a single target with                            -     = -25.4 mm.



</P>
<P><PB REF="00000060.tif" SEQ="00000060" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="625" N="00000060">
bo        soS      1.00      1. 50,0        2'.50    3'.00     3         4. 00       so      5: S00,         LIO        S        10
MICROSECONDS
Figi-re  23.  Computer simulation of the effect of scattering on a single target withd                                        5.    m



</P>
<P><PB REF="00000061.tif" SEQ="00000061" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="695" N="00000061">
al
SS
SM
F
a
8.00.50       1.00      1.50     2.00      2.50      3.00      3.50'1.00'1.50      5.00      5.50     6.00      8.50      7.00
MICROSECONDS
Figure 24. Computer simulation of the effect of scattering on a single target with a1  d = +127 rrmn.



</P>
<P><PB REF="00000062.tif" SEQ="00000062" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="582" N="00000062">
dr
io        io      1.00      1.50      2.00o     2.50     3.00      3.50               14. 0,0   S.00               1.0.O.0
MIICROSE~CONDS
Figure 25.  Computer simulation of the effect of scattering on a single target.   MOf                               xprf)



</P>
<P><PB REF="00000063.tif" SEQ="00000063" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="884" N="00000063">
and the reference function. As Figures 20-25 show, uncompensated scattering
lowers the maximum SNR but maintains the full resolution capability of the
system.
In order to calculate M2(f), it is necessary to determine B1, and B2.
Mason (1950, page 419) presents some experimental data for aluminum with
an average grain size of 0.25 mm that is well-fitted by the following values
of Bi and B2.
B1  =  8.86 x 10-3 dB/mm/MHz
B2  =  3.94 x 10-5 dB/mm/MHz                                        (33)
Once M2(f) is known, then equation 58 is used to find the final output
function, S(f). The effect of the scattering medium can be best illustrated
by using a single point target, in which case G(f) = 1 and
S(f)  =  M2(f) + M2(f) cos(2fT) + 1                                 (34)
S(f) is calculated for various values of d1 - d over a frequency range
approximately the same as the bandwidth of the transducer used in Figure 12.
It is then multiplied by a Hamming window function and inverse Fourier transformed to yield the results shown in Figures 20-24.  When dl - d is negative,
the simulated discontinuity precedes the reference function. When d1 - d is
positive, the reference function precedes the simulated discontinuity. For
all the data presented in Figures 9-18, d1 - d is negative. For comparison
purposes, the target position on the time axis is the same for each figure
even though dl - d varies from figure to figure.
The result of applying this analysis to the case of simulated single
point target is that there is a neglible difference in resolution as a
function of scattering. The main effect of the scattering is to reduce the
maximum SNR.
APPLICATIONS
Figures 9-18 demonstrate that processing the data returned from an
ultrasonic pulse-echo system in the manner described in Section II can increase
the resolution, increase the SNR, and rpovide an output display that is independent of any particular transducer. These three improvements can be used
to increase the information presented in the following ultrasonic ITI situations
56



</P>
<P><PB REF="00000064.tif" SEQ="00000064" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="897" N="00000064">
The increased resolution can distinguish between a rough, extended defect
and porosity. Using the old, low resolution systems, both the porosity and
the extended defect have the same display indications. As the resolution
increases, the rough surface shows many closely spaced targets with no gaps,
while the porosity shows discrete targets separated by distinct gaps.  Thus,
the appearance of gaps can be used to indicate porosity in a high resolution
system.
The increased resolution can also identify defects which are located very
close to a known surface.  In a conventional, low resolution system, the
strong echo from the known surface masks the weak echoes from small targets
located nearby. As the resolution increases, the region over which this
masking occurs diminishes.
High resolution is also potentially important in the testing of spot
welds in thin plates. Existing techniques, both through-transmission and
pulse-echo, rely on the premise that such undesirable factors as low penetration,
porosity, expulsion, and an undersized nugget cause variations in attenuation.
While this premise is true, such tests do not identify the particular type of
defect, nor do they locate the position of the defect within the weld. A
high resolution pulse-echo system has the capability of locating such defects
as expulsion and porosity, and if the acoustic impedance of the fused material
is sufficiently different from the parent material, it can measure penetration
and nugget size as well.
As mentioned in the last part of Section I, this data processing scheme
is intended to be used as part of a scanning system that will display an outline of the defect. In such a system, high resolution is very important since
the determination of the strength of a material containing a discontinuity is
dependent on the orientation of the discontinuity and whether it is large or
small, smooth or rough, spherical or crack-like, etc.
A schematic diagram of a one-dimensional scanning system is shown in
Figure 26. The transducer is placed in position A, and data are collected
and stored exactly as for the data in Figures 8-17. The transducer is then
placed in positions B, C, D, etc., and the data are collected and processed
as in position A. After the scan has been completed, a high resolution (both
lateral and longitudinal) "A" scan for position C, for example, can be formed
by point-by-point addition of the data from positions A-E. The computer
addition of the data from positions A-E simulates the physical addition of
the signal that would occur in a single transducer whose diameter is equal to
the length of the line segment AE.  Thus, high lateral resolution is achieved
by computer addition, and high longitudinal resolution is achieved by computer
deconvolution.
The lateral resolution can be further increased by the addition of more
positions, as long as the transducer beamwidth is adequate to insonify the
57



</P>
<P><PB REF="00000065.tif" SEQ="00000065" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="851" N="00000065">
A  B  C   D  E  F  G
Figure 26.  Schematic diagram of a phased array scanning system.  Points
A through G represent the same transducer at various positions during
the scan.



</P>
<P><PB REF="00000066.tif" SEQ="00000066" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="888" N="00000066">
target. This fact has the interesting consequences that the lateral resolution
is increased by using a smaller diameter transducer, and that the resolution
of the processed data will be independent of the target range.  Since the
transducer beamwidth is inversely proportional to its diameter, a small
transducer will have a large beamwidth.  If the target is located directly
under position C, a large beamwidth transducer will allow positions D, E, F,
etc., to insonify and to collect data from this target, and the more positions
that insonify a given target, the higher the lateral resolution. The rangeindependent resolution is due to the fact that the resolving power is directly
proportional to the equivalent transducer diameter and inversely proportional
to the range while the equivalent transducer diameter is directly proportional
to the range.  Thus the range term cancels out.
From just this short description of a possible scanning system, it can
be seen that it has many inherent advantages over the conventional amplitude
comparison technique of defect characterization. Besides having advantageous
characteristics of its own, the data processing system that has been described
produces data in a form ideally suited for the realization of this scanning
technique.  As equation (12) in Section II shows, the data processing output
contains both the amplitude and phase information present in the reflectivity
function.  Without both amplitude and phase information, it is impossible to
achieve the full benefits of the scanning system.
The SNR improvement offered by the data processing system is also an
important part of the proposed scanning system. Since small diameter transducers will be used in order to increase the lateral resolution of the scanning
system.  The unprocessed SNR at any given position in Figure 26 will be decreased
accordingly.  The signal averaging implemented by the data processing system
increases the SNR before the array processing and prevents the small diameter
transducer from adversely affecting the final SNR. The array processing used
in the scanning system also decreases the noise due to the random grain
structure. Again, this noise is decreased by the signal averaging performed
during the addition of the data from the many different transducer positions.
The SNR improvement demonstrated in Figure 17 can be very important in
the application of ultrasonic NDT to such materials as titanium billets and
centrifugally-cast pipes.  As Sattler (1969) shows, the SNR of a conventional
pulse-echo system used to test titanium billets is very low.  He states that
"A" 10-MHz test frequency proved unfeasible because of high attenuation and
low signal return."  Using the data processing system, the SNR could be improved
enough to enable the testing to proceed at the 10-MHz frequency. Without
data processing, the tests would have to be performed at lower frequencies
with an attendant loss in resolution. In the case of centrifugally-cast pipe,
the microstructure is very large and preferentially oriented along the lines
of force present during the casting process.  This microstructure scatters a
large proportion of the ultrasonic energy and makes it practically impossible
to use conventional ultrasonic NLDT.  The SNR improvement demonstrated in
59



</P>
<P><PB REF="00000067.tif" SEQ="00000067" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="893" N="00000067">
Figure 17 opens up the possibility that defects that are large with respect
to the microstructure may be located by ultrasonic techniques even in
centrifugally-cast structures.
The third advantageous characteristic of the data processing system,
transducer independence, makes display interpretation very easy. Since the
ringing response of the transducer is completely removed, the processed
display has all the advantages of a rectified and filtered display.  A single
target is represented by a single, unipolar indication. The display is also
independent of normal transducer aging and replacement.  This feature is very
important when test results are compared for change over a period of years.
In a conventional system, the transducer may age or need to be replaced, and
when the new test data are compared to the old, apparent changes may be due
to the different transducer rather than to discontinuity growth. If the data
from both the new and the old tests are deconvolved, the effect of the transducer on the test is negligible and any changes in test data can be confidently
attributed to discontinuity growth.
This data processing system can also be applied to shear wave and surface
wave testing. The SNR enhancement and frequency dependent media compensation
are especially important properties for use in shear wave systems since the
lower velocity of shear waves produces a smaller wavelength that is more
easily scattered and attenuated by the material microstructure.
The application of this processing scheme to "pitch-catch" and throughtransmission systems that use two transducers is limited by the complexity
of the necessary deconvolution filter; no simple amplitude- only filter can
be formed.  The SNR enhancement procedure could be applied separately, in
such cases, without any deconvolution, if it is necessary to improve the
sensitivity of the system. But this approach should be considered only after
rejecting any alternative techniques.
60



</P>
<P><PB REF="00000068.tif" SEQ="00000068" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="873" N="00000068">
VI.  CO NCLUSIONS
A new approach to ultrasonic pulse-echo!NDT has been presented.  it
utilizes computer data processing of the information in the echo in conjunction with a computerized scanning system  to  form a visual display containing
an outline of the hidden discontinuity. At present, a large amount of ultrasonic pulse-echo NDT information is being discarded because minimal use is
made of the correlation of the information returned from adjacent transducer
echoes in time and from adjacent transducer positions in space. Experimental
evidence has been presented to demonstrate that the information inherent in
the time sequence of transducer echoes can be processed by computerized signal
averaging to provide a 28 dB increase in the SUR over that available from a
conventional system. This experimental evidence also demonstrates that computer processing of pulse-echo data can increase the longitudinal resolution
by a factor of five and can present a display that is virtually independent
of the physical characteristics of the particular transducer used to collect
the data.
A mathematical model of this system provides an especially simple form
(amplitude only) for the deconvolutlon filter used to increase the longitudinal resolution.  This model also provides a formula expressing the SIRm of the
output in terms of the SNR of the input and the system parameters, such as the
number of samples and the filter bandwidths. An analysis of this model also
shows that, by the proper use of a reference function required in the data
acquisition stage, the effect of frequency-dependent scattering in the medium
can be partially removed.  A compute- simulation, designed to sho~w the remaining residual effect of frequency-dependent scattering in the medium demonstrates that the uncompensated scattering causes a negligible decrease in the
resolution and a slight decrease in the maximum SNR.
Due to the presence of high frequency information in the ravw data, digital data cannot be acquired directly from the ultrasonic echo.  To circumvent
this problem, a data acquisition system which utilizes a holographic approach
preserving both the amplitude and the phase information present in the ultrasonic echo is developed. Since the digital data preserves both the phase and
amplitude of the ultrasonic echo, it can be stored and then recombined at the
completion of a scanning procedure to form a synthetic array of transducers.
Such an array processor should increase the lateral resolution of the transducer and increase the SNR for noise generated by the material microstructure
in the neighborhood of the defect target.
The application of this data processing system to current problems in
ultrasonic pulse-echo ITDT is also discussed.  When applied to immersion testing,
the system  is capable of revealing discontinuities lying closer to the front
surface of a part than can be detected with a conventional system.  The
hi



</P>
<P><PB REF="00000069.tif" SEQ="00000069" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="886" N="00000069">
improved resolution makes it easier t.o differentiate between porosity and a
large extended defect. The transducer-independent visual display is able to
compare test results taken over a time period in which it is possible that the
transducer may have aged or been replaced. And finally, the improved SNR increases the sensitivities of ultrasonic pulse-echo testing in materials such
as cast titanium and to centrifugally cast pipe where the grain structure
causes excessive scattering of the ultrasonic pulse.



</P>
<P><PB REF="00000070.tif" SEQ="00000070" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="880" N="00000070">
VII. FUTURE WORK
The next logical step in the development of this data processing system
is to construct an array processor. Initially, a few simple targets could be
scanned from a plane surface, and the data from each transducer position could
be summed and displayed in order to establish the lateral resolution.  At this
stage, close attention should be paid to the effect of the contact conditions
between the transducer and the test sample.  After this experiment, the scanning system can be generalized to surfaces with arbitrary curvature and to
complex targets.
The data acquisition system used in this research was necessary in order
to accurately measure and record the high frequency data in the ultrasonic
echo.  If sample-and-hold modules with aperture times on the order of 0.5 ns
or less become available, then a much simpler and cheaper data acquisition
system is possible.  Recent developments in commercial sample-and-hold modules
indicate the possibility of such high speed units. In this case, the data
would be collected directly from the ultrasonic echo and no reference function
would be necessary. If the testing material has little attenuation, a single
deconvolution filter can be constructed and applied to all the data.  If the
material is highly attenuating, then a different filter can be constructed for
each distance interval, and in this manner, the effects due to scattering and
attenuation can be minimized. The latter possibility, while a little more
complex than the former, is scarcely more time consuming when implemented on
a digital minicomputer.
For the purposes of this research all the digital data processing was
done on a large, central computer. The necessary processing, though, is quite
simple, and commercially available minicomputers with 8K storage are capable
of handling the necessary data acquisition and processing.  A system designed
around such a minicomputer could be made portable with no loss of performance
and a large gain in convenience.
And finally, with the development of an array processor, the problem of
information display must be considered.  "A" scan, "B" scan, and "C" scan
displays can all be formed by the array processor, but there are many advantages to be gained by considering cross-sectional and isometric projection
displays. A display, or displays, should be chosen to optimize the human interpretation of such large amounts of data.



</P>
<P><PB REF="00000071.tif" SEQ="00000071" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="885" N="00000071">
REFERENCES
Cox, C. W. and Renken, C. J., 1970. "The application of signal-processing
techniques to signals from electromagnetic test systems."  Materials
Evaluation, Vol. 28, No. 8, p. 173.
Davenport, W. B. and Root, W. L., 1958.  Random Signals and Noise, McGraw-Hill
Book Company, Inc., New York.
Gold, B. and Rader, C. M, 1969.  Digital Processing of Signals, McGraw-Hill
Book Company, Inc., New York.
Kaplan, W. F., 1962.  Operational Methods for Linear Systems, Addison-Wesley,
Inc., Reading, Massachusetts.
Mason, W. P., 1950.  Piezoelectric Crystals and Their Application to Ultrasonics, D. Van Nostrand and Company, New York.
Sattler, F. J., 1969. "Nondestructive testing techniques for titanium
billets."  Technical Report AFML-TR-68-345, TRW, Inc., Cleveland, Ohio.
Serabian, S., 1968. "Implications of the attenuation-produced pulse distortion upon the ultrasonic method of nondestructive testing." Materials
Evaluation, Vol. 26, No. 9, p. 174.
Seydel, James A., 1972.  Computerized Enhancement of' Ultrasonic Nondestructive
Testing Data.  Ph.D. Thesis, The University of Michigan (1972).
Sokolov, S. J., 1941.  "Ultrasonic methods for determining the properties of
heat-treated steel and for determining internal flaws of metal objects."
Zhurnal Tekhnicheskoi Fiziki (printed in Russian), Vol. 11, p. 160.
64



</P>
<P><PB REF="00000072.tif" SEQ="00000072" RES="600dpi" FMT="TIFF6.0" FTR="UNSPEC" CNF="487" N="00000072">
UNIVERSITY OF MICHIGAN
3lllllll1:lllll Illlllllll/IJI!1l 9 1 8;1 3111 871111
3 9015 02826 3187



</P>
</DIV1>
</BODY>
</TEXT>