Work Description

Title: Data for "Impact of Tissue Sample Preparation Methods on Myelin-Sensitive Quantitative MR Imaging" Open Access Deposited

h
Attribute Value
Methodology
  • MRI scans of 14 sheep brain tissue samples were acquired on a 7.0 Tesla NMR/MRI scanner (Varian/Agilent). Various scans were acquired: gradient echo, fast spin echo, multi-echo spin echo, B0 mapping, B1 mapping, inversion recovery, and magnetization transfer scans. Histology (Luxol fast blue stain / Cresyl violet counterstain) images were also acquired.
Description
  • Validation of quantitative MRI (qMRI) parameters with histology is often done with ex vivo fixed tissue samples. Freezing is another common form of tissue preservation, but the effects of freezing and thawing tissue on myelin-sensitive qMRI parameters and their correlation with histology require further analysis. To study the effects of freezing vs. fixing tissue on myelin-sensitive parameters, we conducted myelin water imaging, off-resonance RF saturation magnetization transfer (MT), and selective inversion recovery MT MRI experiments on 14 fresh, thawed, and fixed sheep brain tissue samples to calculate various surrogate measures of myelin content. These measures were compared with luxol fast blue (LFB) histological stain results. 14 sheep brain tissue samples were scanned fresh. Half were frozen and subsequently thawed and scanned again. The second half were fixed and scanned again. Then all samples were fixed and sent for histological processing. This folder contains the MR and histology images collected for this study. All data are stored as mat files. The top level folder sheep_data_paper has subfolders: fresh_thawed: holds the data for the samples scanned fresh and thawed (S1-S7) fresh_fixed: holds the data for the samples scanned fresh and fresh_fixed (S8-S14) Each structure variable (data.mat) holds data for a specific sample and a specific tissue preparation (fresh, thawed, or fixed). The data variable has the following fields: gre: gradient echo scan data (128 x 128 x 5) fse: fast spin echo scan data (256 x 256 x 5) mese: multi-echo spin echo scan data for MWF mapping (128 x 128 x 5 x 104) b0data: gradient echo scan data for B0 mapping (128 x 128 x 5 x 4) b1data: EPI scan data for B1 mapping (32 x 32 x 5 x 9) t1data: EPI inversion recovery data for T1 mapping (64 x 64 x 5 x 21) mtdata: gradient echo data for quantitative MT analysis (128 x 128 x 4 x nf) hist: Luxol fast blue stain / Cresyl violet counterstain (LFB) histology image
Creator
Depositor
Depositor creator
  • false
Contact information
Discipline
Funding agency
  • National Institutes of Health (NIH)
ORSP grant number
  • P30AG072931-03S1
Keyword
Citations to related material
  • Murguia, A., Swanson, S. D., Scheven, U., Jacobson, A., Nielsen, J. F., Fessler, J. A., & Seraji-Bozorgzad, N. (2025). Impact of Tissue Sample Preparation Methods on Myelin-Sensitive Quantitative MR Imaging. bioRxiv, 2025-05.
Resource type
Last modified
  • 06/19/2025
Published
  • 06/19/2025
DOI
  • https://doi.org/10.7302/dp7t-c323
License
To Cite this Work:
Murguia, A., Swanson, S. D., Scheven, U., Jacobson, A., Nielsen, J., Fessler, J. A., Seraji-Bozorgzad, N. (2025). Data for "Impact of Tissue Sample Preparation Methods on Myelin-Sensitive Quantitative MR Imaging" [Data set], University of Michigan - Deep Blue Data. https://doi.org/10.7302/dp7t-c323

Relationships

This work is not a member of any user collections.

Files (Count: 2; Size: 833 MB)

Date written: June 8, 2025
README Author: Amaya Murguia
README Author Email: [email protected]

Data accompanying paper: "Impact of Tissue Sample Preparation Methods on Myelin-Sensitive Quantitative MR Imaging"
Citation: Murguia, A., Swanson, S. D., Scheven, U., Jacobson, A., Nielsen, J. F., Fessler, J. A., & Seraji-Bozorgzad, N. (2025).
Impact of Tissue Sample Preparation Methods on Myelin-Sensitive Quantitative MR Imaging. bioRxiv, 2025-05.
https://www.biorxiv.org/content/10.1101/2025.05.05.652075v1.abstract

Research project description: Validation of quantitative MRI (qMRI) parameters with histology is often done with ex vivo fixed tissue samples.
Freezing is another common form of tissue preservation, but the effects of freezing and thawing tissue on myelin-sensitive qMRI parameters
and their correlation with histology require further analysis. To study the effects of freezing vs. fixing tissue on myelin-sensitive parameters,
we conducted myelin water imaging, off-resonance RF saturation magnetization transfer (MT), and selective inversion recovery MT MRI experiments
on 14 fresh, thawed, and fixed sheep brain tissue samples to calculate various surrogate measures of myelin content. These measures were
compared with luxol fast blue (LFB) histological stain results. 14 sheep brain tissue samples were scanned fresh.
Half were frozen and subsequently thawed and scanned again. The second half were fixed and scanned again.
Then all samples were fixed and sent for histological processing.

Methods: MRI scans of 14 sheep brain tissue samples were acquired on a 7.0 Tesla NMR/MRI scanner (Varian/Agilent).
Various scans were acquired: gradient echo, fast spin echo, multi-echo spin echo, B0 mapping, B1 mapping, inversion recovery,
and magnetization transfer scans. Histology (Luxol fast blue stain / Cresyl violet counterstain) images were also acquired.

The top level folder sheep_data_paper has subfolders:
fresh_thawed: holds the data for the samples scanned fresh and thawed (S1-S7)
fresh_fixed: holds the data for the samples scanned fresh and fresh_fixed (S8-S14)

Each structure variable (data.mat) holds data for a specific sample and a specific tissue preparation (fresh, thawed, or fixed).
The data variable has the following fields:

gre: gradient echo scan data (128 x 128 x 5)
fse: fast spin echo scan data (256 x 256 x 5)
mese: multi-echo spin echo scan data for MWF mapping (128 x 128 x 5 x 104)
b0data: gradient echo scan data for B0 mapping (128 x 128 x 5 x 4)
b1data: EPI scan data for B1 mapping (32 x 32 x 5 x 9)
t1data: EPI inversion recovery data for T1 mapping (64 x 64 x 5 x 21)
mtdata: gradient echo data for quantitative MT analysis (128 x 128 x 4 x nf)
hist: Luxol fast blue stain / Cresyl violet counterstain (LFB) histology image

This folder contains the sheep brain tissue MR and histology images collected for this study.
All data are stored as mat files.

top level folder name: sheep_data_paper
subfolders:
fresh_thawed: holds the data for the samples scanned fresh and thawed (S1-S7)
fresh_fixed: holds the data for the samples scanned fresh and fresh_fixed (S8-S14)

Each structure variable (data.mat) holds data for a specific sample and a specific tissue preparation (fresh, thawed, or fixed).
The data variable has the following fields:

gre: gradient echo scan data (128 x 128 x 5)
fse: fast spin echo scan data (256 x 256 x 5)
mese: multi-echo spin echo scan data for MWF mapping (128 x 128 x 5 x 104)
b0data: gradient echo scan data for B0 mapping (128 x 128 x 5 x 4)
b1data: EPI scan data for B1 mapping (32 x 32 x 5 x 9)
t1data: EPI inversion recovery data for T1 mapping (64 x 64 x 5 x 21)
mtdata: gradient echo data for quantitative MT analysis (128 x 128 x 4 x nf)
hist: Luxol fast blue stain / Cresyl violet counterstain (LFB) histology image

Each of the variables (with MR data) is also a structure variable with the following fields:
image (complex image data)
kspace (complex kspace data, only have for the non-EPI data - gre, fse, mese, b0data, mtdata)
pars (parameters pulled off the scanner)
acquisitiontype (2D or 3D)

hist is also a structure with the following fields
LFB_RGB (uint8 of size 257 x 173 x 3, downsampled LFB image)
LFB_B_minus_R (uint8 of size 257 x 173 with the blue minus the red channel of the downsampled RGB LFB image)

Scan parameters (please see the paper for more details):

Slice plan: 5 slices, interleaved, 2mm thickness, 0mm gap

Non-EPI scan details
gre: TR = 100ms, TE = 4ms, matrix size = 128 x 128, 5 slices
fse: TR = 4000ms, TE (effective) = 40ms, echo spacing = 10ms, segments/echo train length = 32/8, k-zero = 4, averages = 1, matrix size = 256 x 256
mese: TR = 4000ms, TE = 5ms, number of echoes = 104, averages = 2, matrix size = 128 x 128
b0data: TR = 100ms, TE array of 4, 6, 8, 10ms, flip angle = 20 deg, averages = 1, matrix size = 128 x 128
mtdata: 1 slice (not 5 for this scan), TR = 120ms, TE = 3ms, flip angle = 20 deg, averages = 1, matrix size = 128 x 128, 4 power levels (satpwr),
nf off-resonance frequencies (satfrq1) (nf = 25 usually, for a few samples nf = 21)

EPI scan details
b1data: TR = 8000ms, TE = 9.55ms (minimum), flip angle = array of 9 values from 20-180 deg spaced 20 deg apart, averages = 1 (1 shot, 1 repetition), matrix size = 32 x 32
t1data: TR = 8000ms, TR = 36.64ms (minimum), averages = 1 (1 shot, 1 repetition), matrix size = 64 x 64, spin echo, TI array of 21 inversion times (log spaced from 10ms-5ms)

Useful parameter names from pars file (these variable names are the Varian scanner parameter names):
ns: number of slices
te: TE in seconds
tr: TR in seconds
ne: number of echoes
flip1: flip angle in degrees
ti: TI (inversion time) in seconds
resto: resonance transmitter offset frequency in Hz
thk: slice thickness in millimeters
pw: pulse width in microseconds
satfrq1: off-resonance frequency in Hz (for MT data preparation pulse)
satpwr: power level in dB (for MT data preparation pulse)

Not every sample has data for every MR measure; please see Table S1 in the paper.

Download All Files (To download individual files, select them in the “Files” panel above)

Best for data sets < 3 GB. Downloads all files plus metadata into a zip file.



Best for data sets > 3 GB. Globus is the platform Deep Blue Data uses to make large data sets available.   More about Globus

Remediation of Harmful Language

The University of Michigan Library aims to describe its collections in a way that respects the people and communities who create, use, and are represented in them. We encourage you to contact us anonymously if you encounter harmful or problematic language in catalog records or finding aids. More information about our policies and practices is available at Remediation of Harmful Language.