Show simple item record

Validation of a robust method for quantification of three-dimensional growth of the thoracic aorta using deformable image registration

dc.contributor.authorBian, Zhangxing
dc.contributor.authorZhong, Jiayang
dc.contributor.authorDominic, Jeffrey
dc.contributor.authorChristensen, Gary E.
dc.contributor.authorHatt, Charles R.
dc.contributor.authorBurris, Nicholas S.
dc.date.accessioned2022-05-06T17:27:17Z
dc.date.available2023-05-06 13:27:13en
dc.date.available2022-05-06T17:27:17Z
dc.date.issued2022-04
dc.identifier.citationBian, Zhangxing; Zhong, Jiayang; Dominic, Jeffrey; Christensen, Gary E.; Hatt, Charles R.; Burris, Nicholas S. (2022). "Validation of a robust method for quantification of three-dimensional growth of the thoracic aorta using deformable image registration." Medical Physics 49(4): 2514-2530.
dc.identifier.issn0094-2405
dc.identifier.issn2473-4209
dc.identifier.urihttps://hdl.handle.net/2027.42/172287
dc.description.abstractPurposeAccurate assessment of thoracic aortic aneurysm (TAA) growth is important for appropriate clinical management. Maximal aortic diameter is the primary metric that is used to assess growth, but it suffers from substantial measurement variability. A recently proposed technique, termed vascular deformation mapping (VDM), is able to quantify three-dimensional aortic growth using clinical computed tomography angiography (CTA) data using an approach based on deformable image registration (DIR). However, the accuracy and robustness of VDM remains undefined given the lack of ground truth from clinical CTA data, and, furthermore, the performance of VDM relative to standard manual diameter measurements is unknown.MethodsTo evaluate the performance of the VDM pipeline for quantifying aortic growth, we developed a novel and systematic evaluation process to generate 76 unique synthetic CTA growth phantoms (based on 10 unique cases) with variable degrees and locations of aortic wall deformation. Aortic deformation was quantified using two metrics: area ratio (AR), defined as the ratio of surface area in triangular mesh elements and the magnitude of deformation in the normal direction (DiN) relative to the aortic surface. Using these phantoms, we further investigated the effects on VDM’s measurement accuracy resulting from factors that influence the quality of clinical CTA data such as respiratory translations, slice thickness, and image noise. Lastly, we compare the measurement error of VDM TAA growth assessments against two expert raters performing standard diameter measurements of synthetic phantom images.ResultsAcross our population of 76 synthetic growth phantoms, the median absolute error was 0.063 (IQR: 0.073–0.054) for AR and 0.181 mm (interquartile range [IQR]: 0.214–0.143 mm) for DiN. Median relative error was 1.4% for AR and 3.3%$3.3%$ for DiN at the highest tested noise level (contrast-to-noise ratio [CNR] = 2.66). Error in VDM output increased with slice thickness, with the highest median relative error of 1.5% for AR and 4.1% for DiN at a slice thickness of 2.0 mm. Respiratory motion of the aorta resulted in maximal absolute error of 3% AR and 0.6 mm in DiN, but bulk translations in aortic position had a very small effect on measured AR and DiN values (relative errors <1%$&lt; 1%$). VDM-derived measurements of magnitude and location of maximal diameter change demonstrated significantly high accuracy and lower variability compared to two expert manual raters (p<0.03$p&lt;0.03$ across all comparisons).ConclusionsVDM yields an accurate, three-dimensional assessment of aortic growth in TAA patients and is robust to factors such as image noise, respiration-induced translations, and differences in patient position. Further, VDM significantly outperformed two expert manual raters in assessing the magnitude and location of aortic growth despite optimized experimental measurement conditions. These results support validation of the VDM technique for accurate quantification of aortic growth in patients and highlight several important advantages over diameter measurements.
dc.publisherWiley Periodicals, Inc.
dc.subject.othervascular deformation mapping
dc.subject.otherdeformable registration
dc.subject.otherthoracic aortic aneurysm
dc.titleValidation of a robust method for quantification of three-dimensional growth of the thoracic aorta using deformable image registration
dc.typeArticle
dc.rights.robotsIndexNoFollow
dc.subject.hlbsecondlevelMedicine (General)
dc.subject.hlbtoplevelHealth Sciences
dc.description.peerreviewedPeer Reviewed
dc.description.bitstreamurlhttp://deepblue.lib.umich.edu/bitstream/2027.42/172287/1/mp15496.pdf
dc.description.bitstreamurlhttp://deepblue.lib.umich.edu/bitstream/2027.42/172287/2/mp15496_am.pdf
dc.identifier.doi10.1002/mp.15496
dc.identifier.sourceMedical Physics
dc.identifier.citedreferenceSuh G-Y, Beygui RE, Fleischmann D, Cheng CP. Aortic arch vessel geometries and deformations in patients with thoracic aortic aneurysms and dissections. J Vasc Interv Radiol. 2014; 25: 1903 - 1911.
dc.identifier.citedreferenceHiratzka LF, et al. ACCF/AHA/AATS/ACR/ASA/SCA/SCAI/SIR/STS/SVM guidelines for the diagnosis and management of patients with thoracic aortic disease: executive summary. Circulation. 2010; 121: 1544.
dc.identifier.citedreferenceGuo MH, Appoo JJ, Saczkowski R, et al. Association of mortality and acute aortic events with ascending aortic aneurysm: a systematic review and meta-analysis. JAMA Network Open. 2018; 1: e181281 - e181281.
dc.identifier.citedreferenceQuint LE, Liu PS, Booher AM, Watcharotone K, Myles JD. Proximal thoracic aortic diameter measurements at CT: repeatability and reproducibility according to measurement method. Int J Cardiovasc Imaging. 2013; 29: 479 - 488.
dc.identifier.citedreferenceBurris N, Bian Z, Dominic J, et al. Vascular deformation mapping for CT surveillance of thoracic aortic aneurysm growth. Radiology. 2022: 302: 218 - 225. https://pubs.rsna.org/doi/10.1148/radiol.2021210658.
dc.identifier.citedreferenceBurris NS, Hoff BA, Patel HJ, Kazerooni EA, Ross BD. Three-dimensional growth analysis of thoracic aortic aneurysm with vascular deformation mapping. Circulation: Cardiovascular Imaging. 2018; 11: e008045.
dc.identifier.citedreferenceKlein S, Staring M, Pluim JP. Evaluation of optimization methods for nonrigid medical image registration using mutual information and B-splines. IEEE Trans Image Process. 2007; 16: 2879 - 2890.
dc.identifier.citedreferenceRueckert D, Sonoda LI, Hayes C, Hill DL, Leach MO, Hawkes DJ. Nonrigid registration using free-form deformations: application to breast MR images. IEEE Trans Med Imaging. 1999; 18: 712 - 721.
dc.identifier.citedreferenceStaring M, Klein S, Pluim JP. A rigidity penalty term for nonrigid registration. Med Phys. 2007; 34: 4098 - 4108.
dc.identifier.citedreferenceThévenaz P, Unser M. Optimization of mutual information for multiresolution image registration. IEEE Trans Image Process. 2000; 9: 2083 - 2099.
dc.identifier.citedreferenceMaes F, Collignon A, Vandermeulen D, Marchal G, Suetens P. Multimodality image registration by maximization of mutual information. IEEE Trans Med Imaging. 1997; 16: 187 - 198.
dc.identifier.citedreferenceKlein S, Staring M, Murphy K, Viergever MA, Pluim JP. Elastix: a toolbox for intensity-based medical image registration. IEEE Trans Med Imaging. 2009; 29: 196 - 205.
dc.identifier.citedreferenceLorensen WE, Cline HE. Marching cubes: A high resolution 3D surface construction algorithm. ACM SIGGRAPH Comput Graph. 1987; 21: 163 - 169.
dc.identifier.citedreferenceIzzo R, Steinman D, Manini S, Antiga L. The vascular modeling toolkit: a Python library for the analysis of tubular structures in medical images. Journal of Open Source Software. 2018; 3: 745.
dc.identifier.citedreferenceLu TLC, Rizzo E, Marques-Vidal PM, Segesser LKV, Dehmeshki J, Qanadli SD. Variability of ascending aorta diameter measurements as assessed with electrocardiography-gated multidetector computerized tomography and computer assisted diagnosis software. Interact Cardiovasc Thorac Surg. 2010; 10: 217 - 221.
dc.identifier.citedreferenceGao X, Boccalini S, Kitslaar PH, et al. A novel software tool for semi-automatic quantification of thoracic aorta dilatation on baseline and follow-up computed tomography angiography. Int J Cardiovasc Imaging. 2019; 35: 711 - 723.
dc.identifier.citedreferenceSubramaniam DR, Stoddard WA, Mortensen KH, et al. Continuous measurement of aortic dimensions in Turner syndrome: a cardiovascular magnetic resonance study. J Cardiovasc Magn Reson. 2017; 19: 1 - 17.
dc.identifier.citedreferenceKalra MK, Maher MM, Toth TL, et al. Techniques and applications of automatic tube current modulation for CT. Radiology. 2004; 233: 649 - 657.
dc.identifier.citedreferencePasta S, Agnese V, Di Giuseppe M, et al. In vivo strain analysis of dilated ascending thoracic aorta by ECG-gated CT angiographic imaging. Ann Biomed Eng. 2017; 45: 2911 - 2920.
dc.identifier.citedreferenceAsch FM, et al. The need for standardized methods for measuring the aorta: multimodality core lab experience from the GenTAC registry. JACC Cardiovasc Imaging. 2016; 9: 219 - 226.
dc.working.doiNOen
dc.owningcollnameInterdisciplinary and Peer-Reviewed


Files in this item

Show simple item record

Remediation of Harmful Language

The University of Michigan Library aims to describe library materials in a way that respects the people and communities who create, use, and are represented in our collections. Report harmful or offensive language in catalog records, finding aids, or elsewhere in our collections anonymously through our metadata feedback form. More information at Remediation of Harmful Language.

Accessibility

If you are unable to use this file in its current format, please select the Contact Us link and we can modify it to make it more accessible to you.