Gamma ray image characteristics in sodium iodide slabs
dc.contributor.author | Knoll, Glenn F. | en_US |
dc.date.accessioned | 2006-04-17T15:19:19Z | |
dc.date.available | 2006-04-17T15:19:19Z | |
dc.date.issued | 1969-06-01 | en_US |
dc.identifier.citation | Knoll, G. F. (1969/06/01)."Gamma ray image characteristics in sodium iodide slabs." Nuclear Instruments and Methods 71(2): 163-169. <http://hdl.handle.net/2027.42/32953> | en_US |
dc.identifier.uri | http://www.sciencedirect.com/science/article/B73DN-4718B7V-40/2/75a4333ab6fc2f16858bd0eb7c00e6bd | en_US |
dc.identifier.uri | https://hdl.handle.net/2027.42/32953 | |
dc.description.abstract | A Monte Carlo calculation has been made of the effects of detector scattering on the gamma ray imaging properties of sodium iodide slabs. All interactions within the detector were assumed to contribute to the image. A detailed accounting was kept of the radial distribution away from the point of incidence of both the number of interactions and the deposited energy. Data are presented for slabs of thickness ranging from and over the energy interval from 140-999 keV. | en_US |
dc.format.extent | 407300 bytes | |
dc.format.extent | 3118 bytes | |
dc.format.mimetype | application/pdf | |
dc.format.mimetype | text/plain | |
dc.language.iso | en_US | |
dc.publisher | Elsevier | en_US |
dc.title | Gamma ray image characteristics in sodium iodide slabs | en_US |
dc.type | Article | en_US |
dc.rights.robots | IndexNoFollow | en_US |
dc.subject.hlbsecondlevel | Physics | en_US |
dc.subject.hlbsecondlevel | Nuclear Engineering and Radiological Sciences | en_US |
dc.subject.hlbtoplevel | Science | en_US |
dc.subject.hlbtoplevel | Engineering | en_US |
dc.description.peerreviewed | Peer Reviewed | en_US |
dc.contributor.affiliationum | University of Michigan, Ann Arbor, Michigan, U.S.A. | en_US |
dc.description.bitstreamurl | http://deepblue.lib.umich.edu/bitstream/2027.42/32953/1/0000336.pdf | en_US |
dc.identifier.doi | http://dx.doi.org/10.1016/0029-554X(69)90007-X | en_US |
dc.identifier.source | Nuclear Instruments and Methods | en_US |
dc.owningcollname | Interdisciplinary and Peer-Reviewed |
Files in this item
Remediation of Harmful Language
The University of Michigan Library aims to describe library materials in a way that respects the people and communities who create, use, and are represented in our collections. Report harmful or offensive language in catalog records, finding aids, or elsewhere in our collections anonymously through our metadata feedback form. More information at Remediation of Harmful Language.