Show simple item record

Thresholds for transesophageal atrial pacing

dc.contributor.authorDick, Macdonald IIen_US
dc.contributor.authorCampbell, Robert M.en_US
dc.contributor.authorJenkins, Janice M.en_US
dc.date.accessioned2006-04-28T16:46:06Z
dc.date.available2006-04-28T16:46:06Z
dc.date.issued1984en_US
dc.identifier.citationDick, MacDonald; Campbell, Robert M.; Jenkins, Janice M. (1984)."Thresholds for transesophageal atrial pacing." Catheterization and Cardiovascular Diagnosis 10(5): 507-513. <http://hdl.handle.net/2027.42/38211>en_US
dc.identifier.issn0098-6569en_US
dc.identifier.issn1097-0304en_US
dc.identifier.urihttps://hdl.handle.net/2027.42/38211
dc.identifier.urihttp://www.ncbi.nlm.nih.gov/sites/entrez?cmd=retrieve&db=pubmed&list_uids=6518515&dopt=citationen_US
dc.description.abstractTo determine the thresholds for transesophageal atrial capture, as well as factors that may influence the thresholds, we measured the minimal current and pulse width required to pace the atria through transesophageal bipolar lead systems in 12 patients, ranging in age from 1 day to 19 years, during 19 episodes of reentrant supraventricular tachycardia. Depending on the patients' age and size several electrode catheters were used. The protocol called for 1-msec step-wise increments in pulse width. At each pulse width the current was increased by 1 mamp until capture was achieved. The mean minimal pulse width and mean minimal current required for capture were 5.8 msec and 13.6 mamp, respectively. Atrial capture was achieved in 75% of attempts at a pulse width and current equal to or less than 6.5 msec and 17.5 mamp, respectively. No correlation between current and pulse width on the one hand and age, height, weight, or body surface area on the other was detected. Likewise neither electrode type nor existence of structural heart disease influenced the threshold required for capture. We conclude that atrial capture can be readily achieved through transesophageal electrodes and is not influenced by the subject's age or size.en_US
dc.format.extent421070 bytes
dc.format.extent3118 bytes
dc.format.mimetypeapplication/pdf
dc.format.mimetypetext/plain
dc.language.isoen_US
dc.publisherWiley Subscription Services, Inc., A Wiley Companyen_US
dc.subject.otherLife and Medical Sciencesen_US
dc.subject.otherCardiovascular Medicineen_US
dc.titleThresholds for transesophageal atrial pacingen_US
dc.typeArticleen_US
dc.rights.robotsIndexNoFollowen_US
dc.subject.hlbsecondlevelInternal Medicine and Specialtiesen_US
dc.subject.hlbtoplevelHealth Sciencesen_US
dc.description.peerreviewedPeer Revieweden_US
dc.contributor.affiliationumC.S. Mott Children's Hospital and the University of Michigan, Ann Arbor ; Division of Pediatric Cardiology, F1123 C.S. Mott Children's Hospital, University of Michigan, 1405 East Ann Street, Ann Arbor, MI 48109-0010en_US
dc.contributor.affiliationumC.S. Mott Children's Hospital and the University of Michigan, Ann Arboren_US
dc.contributor.affiliationumC.S. Mott Children's Hospital and the University of Michigan, Ann Arboren_US
dc.identifier.pmid6518515en_US
dc.description.bitstreamurlhttp://deepblue.lib.umich.edu/bitstream/2027.42/38211/1/1810100515_ftp.pdfen_US
dc.identifier.doihttp://dx.doi.org/10.1002/ccd.1810100515en_US
dc.identifier.sourceCatheterization and Cardiovascular Diagnosisen_US
dc.owningcollnameInterdisciplinary and Peer-Reviewed


Files in this item

Show simple item record

Remediation of Harmful Language

The University of Michigan Library aims to describe library materials in a way that respects the people and communities who create, use, and are represented in our collections. Report harmful or offensive language in catalog records, finding aids, or elsewhere in our collections anonymously through our metadata feedback form. More information at Remediation of Harmful Language.

Accessibility

If you are unable to use this file in its current format, please select the Contact Us link and we can modify it to make it more accessible to you.