Coronary artery dissection caused by exit of the guidewire through the distal perfusion sidehole of an auto-perfusion angioplasty balloon catheter
dc.contributor.author | Werns, Steven W. | en_US |
dc.contributor.author | Bates, Eric R. | en_US |
dc.date.accessioned | 2006-04-28T16:46:37Z | |
dc.date.available | 2006-04-28T16:46:37Z | |
dc.date.issued | 1994-09 | en_US |
dc.identifier.citation | Werns, Steven W.; Bates, Eric R. (1994)."Coronary artery dissection caused by exit of the guidewire through the distal perfusion sidehole of an auto-perfusion angioplasty balloon catheter." Catheterization and Cardiovascular Diagnosis 33(1): 32-35. <http://hdl.handle.net/2027.42/38219> | en_US |
dc.identifier.issn | 0098-6569 | en_US |
dc.identifier.issn | 1097-0304 | en_US |
dc.identifier.uri | https://hdl.handle.net/2027.42/38219 | |
dc.identifier.uri | http://www.ncbi.nlm.nih.gov/sites/entrez?cmd=retrieve&db=pubmed&list_uids=8001099&dopt=citation | en_US |
dc.description.abstract | A previously unrecognized cause of coronary artery dissection is reported. A 67-year-old woman underwent angioplasty of the right coronary artery using an autoperfusion balloon catheter. Dissection occurred because the balloon catheter was advanced while the guidewire exited from one of the distal perfusion sideholes. © Wiley-Liss, Inc. | en_US |
dc.format.extent | 343400 bytes | |
dc.format.extent | 3118 bytes | |
dc.format.mimetype | application/pdf | |
dc.format.mimetype | text/plain | |
dc.language.iso | en_US | |
dc.publisher | Wiley Subscription Services, Inc., A Wiley Company | en_US |
dc.subject.other | Life and Medical Sciences | en_US |
dc.subject.other | Cardiovascular Medicine | en_US |
dc.title | Coronary artery dissection caused by exit of the guidewire through the distal perfusion sidehole of an auto-perfusion angioplasty balloon catheter | en_US |
dc.type | Article | en_US |
dc.rights.robots | IndexNoFollow | en_US |
dc.subject.hlbsecondlevel | Internal Medicine and Specialties | en_US |
dc.subject.hlbtoplevel | Health Sciences | en_US |
dc.description.peerreviewed | Peer Reviewed | en_US |
dc.contributor.affiliationum | Division of Cardiology, University Hospital, University of Michigan Medical Center, Ann Arbor ; Division of Cardiology, University Hospital, University of Michigan Medical Center, B1-F245 Box 0022, 1500 E. Medical Center Drive, Ann Arbor, MI 48109 | en_US |
dc.contributor.affiliationum | Division of Cardiology, University Hospital, University of Michigan Medical Center, Ann Arbor | en_US |
dc.identifier.pmid | 8001099 | en_US |
dc.description.bitstreamurl | http://deepblue.lib.umich.edu/bitstream/2027.42/38219/1/1810330109_ftp.pdf | en_US |
dc.identifier.doi | http://dx.doi.org/10.1002/ccd.1810330109 | en_US |
dc.identifier.source | Catheterization and Cardiovascular Diagnosis | en_US |
dc.owningcollname | Interdisciplinary and Peer-Reviewed |
Files in this item
Remediation of Harmful Language
The University of Michigan Library aims to describe library materials in a way that respects the people and communities who create, use, and are represented in our collections. Report harmful or offensive language in catalog records, finding aids, or elsewhere in our collections anonymously through our metadata feedback form. More information at Remediation of Harmful Language.
Accessibility
If you are unable to use this file in its current format, please select the Contact Us link and we can modify it to make it more accessible to you.