Show simple item record

Wireless Magnetoelastic Monitoring of Biliary Stents.

dc.contributor.authorGreen, Scott Ryanen_US
dc.date.accessioned2009-09-03T14:48:04Z
dc.date.available2010-10-05T18:27:29Zen_US
dc.date.issued2009en_US
dc.date.submitted2009en_US
dc.identifier.urihttps://hdl.handle.net/2027.42/63744
dc.description.abstractThis thesis describes a wireless magnetoelastic system for monitoring the mass accumulation and viscosity changes that accompany the progressive occlusion of biliary stents. The system comprises four components: a biliary stent, a magnetoelastic sensor, a biasing magnet, and an external interrogation module. Design considerations for each component are highlighted and predictive analytical and finite element tools are developed. Two system generations are detailed. The first-generation system utilizes a 37.5 mm x 2 mm ribbon sensor and neodymium magnets to bias the sensor. The second-generation system comprises a patterned 7.5 mm x 29 mm sensor and a strontium-ferrite-polydimethylsiloxane magnetized layer that conform to the stent topology and curvature. Both generations utilize photochemical machining for batch-patterned stents. A resonant frequency change of 2.8% and 6.5% for each respective generation is observed as the viscosity is varied from 1-10 cP. Resonant frequency response to mass loading is similar for both generations, showing a 40% decrease after applying a mass load of 2.5x the mass of the sensor. Advanced sensor designs are also detailed, including a sensor that can evaluate the sludge distribution along the stent length with a resolution of 5 mm, a sensor with varying feature density resulting in a sensitivity increase of 2x compared to other designs, and a hybrid-ribbon sensor that has a full-scale range of at least 10x the mass of the sensor. Finally, an in situ experiment is conducted with a system implanted into a porcine carcass, demonstrating a wireless range of 5 cm. This work has investigated the application of wireless magnetoelastic resonant sensors to monitoring of biliary stents. Methods for interpreting the sensor response to viscous and viscoelastic loads have been demonstrated. In addition, the magnetoelastic sensor performance has been pushed to new limits in terms of full-scale range and mechanical robustness. Advances include development and preliminary experimental verification of modeling tools for resonant magnetoelastic sensors, tailoring of magnetoelastic sensor geometry for enhanced functionality, and demonstration of photochemical machining as a viable process for batch-patterned sensors and stents. Importantly, this work illustrates the long-term potential for passive, wireless magnetoelastic systems for monitoring progressive occlusion of stents.en_US
dc.format.extent30101230 bytes
dc.format.extent1373 bytes
dc.format.mimetypeapplication/pdf
dc.format.mimetypetext/plain
dc.language.isoen_USen_US
dc.subjectMagnetoelasticen_US
dc.subjectResonant Sensorsen_US
dc.subjectBulk Metal Microstructuresen_US
dc.subjectPhotochemical Machiningen_US
dc.subjectBiliary Stentsen_US
dc.titleWireless Magnetoelastic Monitoring of Biliary Stents.en_US
dc.typeThesisen_US
dc.description.thesisdegreenamePhDen_US
dc.description.thesisdegreedisciplineMechanical Engineeringen_US
dc.description.thesisdegreegrantorUniversity of Michigan, Horace H. Rackham School of Graduate Studiesen_US
dc.contributor.committeememberGianchandani, Yogesh B.en_US
dc.contributor.committeememberAwtar, Shoryaen_US
dc.contributor.committeememberElta, Grace H.en_US
dc.contributor.committeememberKurabayashi, Katsuoen_US
dc.contributor.committeememberWise, Kensall D.en_US
dc.subject.hlbsecondlevelBiomedical Engineeringen_US
dc.subject.hlbsecondlevelElectrical Engineeringen_US
dc.subject.hlbsecondlevelMechanical Engineeringen_US
dc.subject.hlbtoplevelEngineeringen_US
dc.description.bitstreamurlhttp://deepblue.lib.umich.edu/bitstream/2027.42/63744/1/greensr_1.pdf
dc.owningcollnameDissertations and Theses (Ph.D. and Master's)


Files in this item

Show simple item record

Remediation of Harmful Language

The University of Michigan Library aims to describe library materials in a way that respects the people and communities who create, use, and are represented in our collections. Report harmful or offensive language in catalog records, finding aids, or elsewhere in our collections anonymously through our metadata feedback form. More information at Remediation of Harmful Language.

Accessibility

If you are unable to use this file in its current format, please select the Contact Us link and we can modify it to make it more accessible to you.