Pragmatic Image Reconstruction for the MiCES Fully-3D Mouse Imaging PET Scanner
dc.contributor.author | Lee, Kisung | en_US |
dc.contributor.author | Kinahan, Paul E. | en_US |
dc.contributor.author | Fessler, Jeffrey A. | en_US |
dc.contributor.author | Miyaoka, Robert S. | en_US |
dc.contributor.author | Janes, Marie | en_US |
dc.contributor.author | Lewellen, Tom K. | en_US |
dc.date.accessioned | 2011-08-18T18:20:54Z | |
dc.date.available | 2011-08-18T18:20:54Z | |
dc.date.issued | 2003-10-19 | en_US |
dc.identifier.citation | Lee, K.; Kinahan, P.E.; Fessler, J.A.; Miyaoka, R.S.; Lewellen, T.K. (2003). "Pragmatic Image Reconstruction for the MiCES Fully-3D Mouse Imaging PET Scanner." IEEE Nuclear Science Symposium Conference Record 4: 2566-2570. <http://hdl.handle.net/2027.42/85855> | en_US |
dc.identifier.uri | https://hdl.handle.net/2027.42/85855 | |
dc.description.abstract | We present a pragmatic approach to image reconstruction for data from the MiCES fully-3D mouse imaging PET scanner under construction at the University of Washington. Our approach is modeled on fully-3D image reconstruction used in clinical PET scanners, which is based on Fourier rebinning (FORE) followed by 2D iterative image reconstruction. The use of iterative methods allows modeling the effects of statistical noise and attenuation etc., while FORE accelerates the reconstruction process by reducing the fully-3D data to a stacked set of independent 2D sinograms. Preliminary investigations have indicated that nonstationary detector point-spread response effects, which are ignored for clinical imaging, significantly impact image quality for the MiCES scanner geometry. To model the effect of nonstationary detector point spread response, we have added a factorized system matrix to the ASPIRE reconstruction library. The current implementation uses FORE+AWOSEM followed by postreconstruction 3D Gaussian smoothing. The results indicate that the proposed approach produces a dramatic improvement in resolution without undue increases in noise. | en_US |
dc.publisher | IEEE | en_US |
dc.title | Pragmatic Image Reconstruction for the MiCES Fully-3D Mouse Imaging PET Scanner | en_US |
dc.type | article | en_US |
dc.subject.hlbsecondlevel | Biomedical Engineering | en_US |
dc.subject.hlbtoplevel | Engineering | en_US |
dc.description.peerreviewed | Peer Reviewed | en_US |
dc.description.bitstreamurl | http://deepblue.lib.umich.edu/bitstream/2027.42/85855/1/Fessler189.pdf | |
dc.identifier.doi | 10.1109/NSSMIC.2003.1352414 | en_US |
dc.identifier.source | IEEE Nuclear Science Symposium Conference Record | en_US |
dc.owningcollname | Electrical Engineering and Computer Science, Department of (EECS) |
Files in this item
Remediation of Harmful Language
The University of Michigan Library aims to describe library materials in a way that respects the people and communities who create, use, and are represented in our collections. Report harmful or offensive language in catalog records, finding aids, or elsewhere in our collections anonymously through our metadata feedback form. More information at Remediation of Harmful Language.
Accessibility
If you are unable to use this file in its current format, please select the Contact Us link and we can modify it to make it more accessible to you.