Incorporation of System Resolution Compensation (RC) in the Ordered-Subset Transmission (OSTR) Algorithm for Transmission Imaging in SPECT

dc.contributor.authorFeng, Bingen_US
dc.contributor.authorFessler, Jeffrey A.en_US
dc.contributor.authorKing, Michael A.en_US
dc.date.accessioned2011-08-18T18:21:24Z
dc.date.available2011-08-18T18:21:24Z
dc.date.issued2006-06-26en_US
dc.identifier.citationFeng, B.; Fessler, J. A.; King, M. A. (2006). "Incorporation of System Resolution Compensation (RC) in the Ordered-Subset Transmission (OSTR) Algorithm for Transmission Imaging in SPECT." IEEE Transactions on Medical Imaging 25(7): 941-949. <http://hdl.handle.net/2027.42/86030>en_US
dc.identifier.issn0278-0062en_US
dc.identifier.urihttps://hdl.handle.net/2027.42/86030
dc.description.abstractIn order to reconstruct attenuation maps with improved spatial resolution and quantitative accuracy, we developed an approximate method of incorporating system resolution compensation (RC) in the ordered-subset transmission (OSTR) algorithm for transmission reconstruction. Our method approximately models the blur caused by the finite intrinsic detector resolution, the nonideal source collimation and detector collimation. We derived the formulation using the optimization transfer principle as in the derivation of the OSTR algorithm. The formulation includes one forward-blur step and one back-blur step, which do not severely slow down reconstruction. The formulation could be applicable to various transmission geometries, such as point-source, line-source, and sheet-source systems. Through computer simulations of the MCAT phantom and transmission measurements of the air-filled Data Spectrum Deluxe single photo emission computed tomography (SPECT) Phantom on a system which employed a cone-beam geometry and a system which employed a scanning-line-source geometry, we showed that incorporation of RC increased spatial resolution and improved the quantitative accuracy of reconstruction. In simulation studies, attenuation maps reconstructed with RC correction improved the quantitative accuracy of emission reconstruction.en_US
dc.publisherIEEEen_US
dc.titleIncorporation of System Resolution Compensation (RC) in the Ordered-Subset Transmission (OSTR) Algorithm for Transmission Imaging in SPECTen_US
dc.typeArticleen_US
dc.subject.hlbsecondlevelBiomedical Engineeringen_US
dc.subject.hlbtoplevelEngineeringen_US
dc.description.peerreviewedPeer Revieweden_US
dc.contributor.affiliationotherDepartment of Radiology, University of Massachusetts Medical School, Worcester, MA 01655 USA.en_US
dc.identifier.pmid16827494en_US
dc.description.bitstreamurlhttp://deepblue.lib.umich.edu/bitstream/2027.42/86030/1/Fessler42.pdf
dc.identifier.doi10.1109/TMI.2006.876151en_US
dc.identifier.sourceIEEE Transactions on Medical Imagingen_US
dc.owningcollnameElectrical Engineering and Computer Science, Department of (EECS)


Files in this item

Show simple item record

Remediation of Harmful Language

The University of Michigan Library aims to describe its collections in a way that respects the people and communities who create, use, and are represented in them. We encourage you to Contact Us anonymously if you encounter harmful or problematic language in catalog records or finding aids. More information about our policies and practices is available at Remediation of Harmful Language.