The contained data comprises what was collected during the characterization of the quad-magnetometer as described in 'Quad-Mag Board for CubeSat Applications'. There are approximately 38 hours of data that compromise a stability test, 10 hours of noise floor testing data, and 10 minutes of sensitivity testing data. Each data file has three-axis measurements from four individual magnetometers over the specified time period at a 65 Hz sampling rate.
Strabel, B. P., Regoli, L. H., Moldwin, M. B., Ojeda, L. V., Shi, Y., Thoma, J. D., Narrett, I. S., Bronner, B., and Pellioni, M.: Quad-Mag board for CubeSat applications, Geosci. Instrum. Method. Data Syst., 11, 375–388, https://doi.org/10.5194/gi-11-375-2022, 2022.
The research analyzed the response of nine PNI RM3100 magnetometers to radiation doses expected during a Europa lander mission. The radiation levels are drawn from the Europa Lander Science Definition Team report ( https://europa.nasa.gov/resources/58/europa-lander-study-2016-report). The sensors were tested up to a total ionization dose (TID) level of 500 kRad.
Regoli, L. H., Moldwin, M. B., Raines, C., Nordheim, T. A., Miller, C. A., Carts, M., and Pozzi, S. A.: Radiation tolerance of the PNI RM3100 magnetometer for a Europa lander mission, Geosci. Instrum. Method. Data Syst., 9, 499–507, https://doi.org/10.5194/gi-9-499-2020, 2020.