Show simple item record

An Optofluidic Lens Biochip and an x-ray Readable Blood Pressure Microsensor: Versatile Tools for in vitro and in vivo Diagnostics.

dc.contributor.authorGulari, Mayurachat Ningen_US
dc.date.accessioned2015-05-14T16:25:22Z
dc.date.availableNO_RESTRICTIONen_US
dc.date.available2015-05-14T16:25:22Z
dc.date.issued2015en_US
dc.date.submitted2015en_US
dc.identifier.urihttps://hdl.handle.net/2027.42/111384
dc.description.abstractThree different microfabricated devices were presented for use in vivo and in vitro diagnostic biomedical applications: an optofluidic-lens biochip, a hand held digital imaging system and an x-ray readable blood pressure sensor for monitoring restenosis. An optofluidic biochip–termed the ‘Microfluidic-based Oil-Immersion Lens’ (mOIL) biochip were designed, fabricated and test for high-resolution imaging of various biological samples. The biochip consists of an array of high refractive index (n = 1.77) sapphire ball lenses sitting on top of an oil-filled microfluidic network of microchambers. The combination of the high optical quality lenses with the immersion oil results in a numerical aperture (NA) of 1.2 which is comparable to the high NA of oil immersion microscope objectives. The biochip can be used as an add-on-module to a stereoscope to improve the resolution from 10 microns down to 0.7 microns. It also has a scalable field of view (FOV) as the total FOV increases linearly with the number of lenses in the biochip (each lens has ~200 microns FOV). By combining the mOIL biochip with a CMOS sensor, a LED light source in 3D printed housing, a compact (40 grams, 4cmx4cmx4cm) high resolution (~0.4 microns) hand held imaging system was developed. The applicability of this system was demonstrated by counting red and white blood cells and imaging fluorescently labelled cells. In blood smear samples, blood cells, sickle cells, and malaria-infected cells were easily identified. To monitor restenosis, an x-ray readable implantable blood pressure sensor was developed. The sensor is based on the use of an x-ray absorbing liquid contained in a microchamber. The microchamber has a flexible membrane that is exposed to blood pressure. When the membrane deflects, the liquid moves into the microfluidic-gauge. The length of the microfluidic-gauge can be measured and consequently the applied pressure exerted on the diaphragm can be calculated. The prototype sensor has dimensions of 1x0.6x10mm and adequate resolution (19mmHg) to detect restenosis in coronary artery stents from a standard chest x-ray. Further improvements of our prototype will open up the possibility of measuring pressure drop in a coronary artery stent in a non-invasively manner.en_US
dc.language.isoen_USen_US
dc.subjectMicrofabricated Microfluidic-based Oil-Immersion Lens (mOIL) biochipen_US
dc.subjectMicrofabricated x-ray readable implantable blood pressure sensoren_US
dc.titleAn Optofluidic Lens Biochip and an x-ray Readable Blood Pressure Microsensor: Versatile Tools for in vitro and in vivo Diagnostics.en_US
dc.typeThesisen_US
dc.description.thesisdegreenamePhDen_US
dc.description.thesisdegreedisciplineMacromolecular Science and Engineeringen_US
dc.description.thesisdegreegrantorUniversity of Michigan, Horace H. Rackham School of Graduate Studiesen_US
dc.contributor.committeememberChronis, Nikolaosen_US
dc.contributor.committeememberKurabayashi, Katsuoen_US
dc.contributor.committeememberRiddell Iv, Jamesen_US
dc.contributor.committeememberMeyhofer, Edgaren_US
dc.contributor.committeememberWise, Kensall D.en_US
dc.subject.hlbsecondlevelBiomedical Engineeringen_US
dc.subject.hlbtoplevelEngineeringen_US
dc.description.bitstreamurlhttp://deepblue.lib.umich.edu/bitstream/2027.42/111384/1/toning_1.pdf
dc.owningcollnameDissertations and Theses (Ph.D. and Master's)


Files in this item

Show simple item record

Remediation of Harmful Language

The University of Michigan Library aims to describe library materials in a way that respects the people and communities who create, use, and are represented in our collections. Report harmful or offensive language in catalog records, finding aids, or elsewhere in our collections anonymously through our metadata feedback form. More information at Remediation of Harmful Language.