Show simple item record

Effects of magnification and zooming on depth perception in digital stereomammography: an observer performance study

dc.contributor.authorChan, Heang-Pingen_US
dc.contributor.authorGoodsitt, Mitchell M.en_US
dc.contributor.authorHadjiiski, Lubomir M.en_US
dc.contributor.authorBailey, Janet E.en_US
dc.contributor.authorKlein, Katherineen_US
dc.contributor.authorDarner, Katie L.en_US
dc.contributor.authorSahiner, Berkmanen_US
dc.date.accessioned2006-12-19T19:04:08Z
dc.date.available2006-12-19T19:04:08Z
dc.date.issued2003-11-21en_US
dc.identifier.citationChan, Heang-Ping; Goodsitt, Mitchell M; Hadjiiski, Lubomir M; Bailey, Janet E; Klein, Katherine; Darner, Katie L; Sahiner, Berkman (2003). "Effects of magnification and zooming on depth perception in digital stereomammography: an observer performance study." Physics in Medicine and Biology. 48(22): 3721-3734. <http://hdl.handle.net/2027.42/48976>en_US
dc.identifier.issn0031-9155en_US
dc.identifier.urihttps://hdl.handle.net/2027.42/48976
dc.identifier.urihttp://www.ncbi.nlm.nih.gov/sites/entrez?cmd=retrieve&db=pubmed&list_uids=14680269&dopt=citationen_US
dc.description.abstractWe are evaluating the application of stereoscopic imaging to digital mammography. In the current study, we investigated the effects of magnification and zooming on depth perception. A modular phantom was designed which contained six layers of 1-mm-thick Lexan plates, each spaced 1 mm apart. Eight to nine small, thin nylon fibrils were pasted on each plate in horizontal or vertical orientations such that they formed 25 crossing fibril pairs in a projected image. The depth separation between each fibril pair ranged from 2 to 10 mm. A change in the order of the Lexan plates changed the depth separation of the two fibrils in a pair. Stereoscopic image pairs of the phantom were acquired with a GE full-field digital mammography system. Three different phantom configurations were imaged. All images were obtained using a Rh target/Rh filter spectrum at 30 kVp tube potential and a ±3° stereo shift angle. Images were acquired in both contact and 1.8X magnification geometry and an exposure range of 4 to 63 mAs was employed. The images were displayed on a Barco monitor driven by a Metheus stereo graphics board and viewed with LCD stereo glasses. Five observers participated in the study. Each observer visually judged whether the vertical fibril was in front of or behind the horizontal fibril in each fibril pair. It was found that the accuracy of depth discrimination increased with increasing fibril depth separation and x-ray exposure. The accuracy was not improved by electronic display zooming of the contact stereo images by 2X. Under conditions of high noise (low mAs) and small depth separation between the fibrils, the observers' depth discrimination ability was significantly better in stereo images acquired with geometric magnification than in images acquired with a contact technique and displayed with or without zooming. Under our experimental conditions, a 2 mm depth discrimination was achieved with over 60% accuracy on contact images with and without zooming, and with over 90% accuracy on magnification images. This study indicates that stereoscopic imaging, especially with magnification, may be useful for visualizing the spatial distribution of microcalcifications in a cluster and for differentiating overlapping tissues from masses on mammograms.en_US
dc.format.extent3118 bytes
dc.format.extent297129 bytes
dc.format.mimetypetext/plain
dc.format.mimetypeapplication/pdf
dc.language.isoen_US
dc.publisherIOP Publishing Ltden_US
dc.titleEffects of magnification and zooming on depth perception in digital stereomammography: an observer performance studyen_US
dc.typeArticleen_US
dc.subject.hlbsecondlevelPhysicsen_US
dc.subject.hlbtoplevelScienceen_US
dc.description.peerreviewedPeer Revieweden_US
dc.contributor.affiliationumDepartment of Radiology, University of Michigan, Ann Arbor, MI 48109, USAen_US
dc.contributor.affiliationumDepartment of Radiology, University of Michigan, Ann Arbor, MI 48109, USAen_US
dc.contributor.affiliationumDepartment of Radiology, University of Michigan, Ann Arbor, MI 48109, USAen_US
dc.contributor.affiliationumDepartment of Radiology, University of Michigan, Ann Arbor, MI 48109, USAen_US
dc.contributor.affiliationumDepartment of Radiology, University of Michigan, Ann Arbor, MI 48109, USAen_US
dc.contributor.affiliationumDepartment of Radiology, University of Michigan, Ann Arbor, MI 48109, USAen_US
dc.contributor.affiliationumDepartment of Radiology, University of Michigan, Ann Arbor, MI 48109, USAen_US
dc.contributor.affiliationumcampusAnn Arboren_US
dc.identifier.pmid14680269en_US
dc.description.bitstreamurlhttp://deepblue.lib.umich.edu/bitstream/2027.42/48976/2/pmb3_22_007.pdfen_US
dc.identifier.doihttp://dx.doi.org/10.1088/0031-9155/48/22/007en_US
dc.identifier.sourcePhysics in Medicine and Biology.en_US
dc.owningcollnameInterdisciplinary and Peer-Reviewed


Files in this item

Show simple item record

Remediation of Harmful Language

The University of Michigan Library aims to describe library materials in a way that respects the people and communities who create, use, and are represented in our collections. Report harmful or offensive language in catalog records, finding aids, or elsewhere in our collections anonymously through our metadata feedback form. More information at Remediation of Harmful Language.

Accessibility

If you are unable to use this file in its current format, please select the Contact Us link and we can modify it to make it more accessible to you.