Work Description

Title: NIH/NCI Co-Clinical Imaging Research Resources Program (CIRP) Diffusion Weighted Imaging Phantom Round Robin Open Access Deposited

h
Attribute Value
Methodology
  • A temperature-controlled phantom provided an absolute reference standard to measure apparent diffusion coefficient (ADC) on pre-clinical MRI scanners using standardized procedures. Spatial uniformity of ADC bias and variance were assessed. Seven institutions participated in the study where diffusion weighted imaging (DWI) datasets were acquired over multiple days on 10 pre-clinical scanners, from 3 vendors at 6 field strengths. Centralized versus site-based analysis was also compared to illustrate incremental variance due to processing workflow.
Description
  • Goals of this work quantify repeatability and reproducibility across multiple NIH/NCI Co-Clinical Imaging Research Resources Program (CIRP) preclinical MRI systems in performing diffusion measurement using a standardized acquisition protocol on a shared phantom containing temperature-controlled medium of known diffusivity. To achieve these goals, the CIRP Image Acquisition Data Processing (IADP) working group (WG) performed a round-robin study of an ice water-based DWI phantom using a detailed phantom preparation procedure and standardized DWI acquisition protocol, with both site- and core-lab generated ADC measurements derived from shared datasets.
Creator
Creator ORCID
Depositor
  • tlchenev@umich.edu
Contact information
Discipline
Funding agency
  • National Institutes of Health (NIH)
ORSP grant number
  • NIH U24 CA237683
Keyword
Date coverage
  • 2021-02-23 to 2022-04-11
Citations to related material
Related items in Deep Blue Documents
Resource type
Last modified
  • 08/01/2023
Published
  • 08/01/2023
Language
DOI
  • https://doi.org/10.7302/zx39-7s47
License
To Cite this Work:
Chenevert, T. L., Malyarenko, D., Amouzandeh, G., Pickup, S., Zhou, R., Manning, H. C., Gamon, S. T., Shoghi, K. I., Quirk, J. D., Sriram, R., Larson, P., Lewis, M. T., Pautler, R. G., Kinahan, P. E., Muzi, M. (2023). NIH/NCI Co-Clinical Imaging Research Resources Program (CIRP) Diffusion Weighted Imaging Phantom Round Robin [Data set], University of Michigan - Deep Blue Data. https://doi.org/10.7302/zx39-7s47

Relationships

This work is not a member of any user collections.

Files (Count: 5; Size: 1.61 GB)

Download All Files (To download individual files, select them in the “Files” panel above)

Total work file size of 1.61 GB may be too large to download directly. Consider using Globus (see below).



Best for data sets > 3 GB. Globus is the platform Deep Blue Data uses to make large data sets available.   More about Globus

Remediation of Harmful Language

The University of Michigan Library aims to describe library materials in a way that respects the people and communities who create, use, and are represented in our collections. Report harmful or offensive language in catalog records, finding aids, or elsewhere in our collections anonymously through our metadata feedback form. More information at Remediation of Harmful Language.